5 | Operation
G 1720
14 / 25
B-H86.0.0C.DB2-1.0
5. In order to save the new parameter value, press and hold the
Function key
for
longer than 1 second.
Representation
Call up menu
Hold
Release
Outcome of an action
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
5.2.5 Configuring parameters of the adjustment menu
Description
The following representation shows the available parameters and various configura-
tion options.
Prerequisites
The
Adjustment
menu is open. See
Adjustment of the measuring input [
}
p. 13]
.
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.
Representation
Parameter
Values
Meaning
Zero point correction
T.OF
0.00
No zero point correction
-5.00 .. 5.00
Zero point correction in °C. and/or at °F -9.00 ..
9.00
Gradient correction of the temperature
T.SL
0.00
No gradient correction
-5.00 .. 5.00
Gradient correction in %
Formula
Zero point correction:
Displayed value = measured value –
T.OF
Gradient correction °C:
Display = (measured value –
T.OF
) * (1 +
T.SL
/ 100)
Gradient correction °F:
Display = (measured value – 32 °F –
T.OF
) * (1 +
T.SL
/ 100) + 32 °F
Example calculation
– Zero point correction
T.OF
to
0.00
– Gradient correction
T.SL
to
0.00
– Display unit
UNIT
to °C
– Display in ice water -0.2 °C
– Display in ice water setpoint
T.OF
= 0.0 °C