E39.0.0X.6C-03 Operating instruction T-Logg 100 and T-Logg 100 E page 4 of 6
BATTERY
:
• Changing display
BAT
/ Value:
The battery of the
T-Logg
is getting discharged soon.
Replace the battery briefly. Even so data logging is still active.
• Constant display
BAT
:
The Logger battery is empty now and data logging has been stopped.
Replace the battery immediately.
ALARM LOW
:
The temperature measured is below the min. alarm limit.
ALARM HIGH
:
The temperature measured has exceeded the max. alarm limit.
ERROR 1
:
The temperature has exceeded the measuring range of the logger.
ERROR 2
:
The temperature has been fallen below the measuring range of the logger.
ERROR 7
:
The
T-Logg
has detected a system fault.
•
Remedy:
Remove battery and wait about 30 minutes. Then insert the battery again.
•
If the error message displayed furthermore, please send the logger to the
manufacturer to repair.
7 Battery service life and recording time
Measuring cycle:
2 sec.
10 sec.
15 min.
5 hours
Recording time:
8.9 hours
44 hours
166 days
Approx. 9 years
Battery service life:
—
Approx. 1 year
Approx. 3 years
—
Please note!
Short measuring cycles as well as frequently measuring data transfer result in a reduction of
the battery service life!
Even if the T-Logg is connected, power consumption is increased. So it’s important to
connect the logger with the USB interface of the PC only as long as necessary!