Inspection and small maintenance for ACS 900
Area
Check or
Clean
Replace if
necessary
Remark
Body
Clean body with a soft towel. Do not use
aggressive cleaning liquid.
Doors / Side panels
Clean the doors and side panels with a soft towel.
Do not use aggressive cleaning liquid.
Filter
Check the filter pads; they can be cleaned. If too
dirty, replace them.
Filter pad
9650.01F.011
Plate
Clean the plate with a soft towel. Do not use
aggressive cleaning liquid.
Rubber discs
Clean the rubber discs with a soft towel. Do not
use aggressive cleaning liquid.
If too used, replace it.
4× Rubber disc (cover)
0322.350.014
1 × Rubber disc (basket holder)
0322.310.030
Contacts of ultrasonic
Check and clean the ultrasonic contacts inside of
the machine (in the whole where the ultrasonic
tank is positioned)
Switch off the device and disconnect the
power cable before doing this.
Ultrasonic tank
Check and clean the contacts on the underside of
ultrasonic tank.
Mechanic
Check if play in the mechanic or large Vibration
during spinning.
Contact service centre