background image

 

Greiner Vibrograf AG 

Mittelstrasse 2 

4900 Langenthal 

Switzerland 

T: +41 62 916 60 80 

F: +41 62 916 60 81 

[email protected] 

www.greinervibrograf.ch 

 

Änderungen vorbehalten 

Seite 2 / 2 

Copyright by Greiner Vibrogaf AG 

 

Repair 

Problem 

Possible cause 

Possible helps 

No function 

 

ACS 900 not switched on 

 

Press the power switch 

 

Power cord not plugged 

 

 Insert cable 

 

Fuse blown 

 

Replace fuse 

 

Building no power 

 

Check power supply 

Error message on display 

 

Mechanical defect 

 

Contact Service Centre 

Ultrasonic is not running 

 

Ultrasound in the program 
not selected 

 

Activate in the program 

 

Ultrasound cannot be 
selected in the program 

 

Contact Service Centre 

Heating is not running 

 

Heating in the program not 
selected 

 

Activate in the program 

 

 Heating defective 

 

Contact Service Centre 

Vibration, mechanical Noise, 
especially during spinning 

 

Play of mechanic 

 

Contact Service Centre 

Plate is not centered 

 

Mechanical defect 

 

Contact Service Centre 

Keyboard does not respond 

 

 Keyboard defective 

 

Contact Service Centre 

Cleaned parts are not clean 

 

 Liquids are too dirty 

 

· Liquids are too old 

 

· Liquids are too long in the 
ACS 900 

 

Check liquids and replace if 
necessary 

Baskets touch glasses 

 

Plate is not centered 

 

Contact Service Centre 

ACS 900 makes noises 

 

Mechanical defect 

 

Contact Service Centre 

 

 

General remarks: 

 

 

Do not use more than 0.55l of liquid for each bath. 

 

The ultrasonic goblet is not waterproof. 

 

Отзывы: