4.8 Performance criteria and limitations
1) PERFORMANCE CRITERIA
A. Mechanical / Optical precision of detection
+/- 0.2 mm
B. Coefficient of determination (reading time 15 min.)
R
2
= 0.958
C. Coefficient of determination (reading time 30 min.)
R
2
= 0.970
D. Automatic temperature conversion
to 18°C. (Manley table)
Accepted range: 15°C - 32°C
E. Acceptable range for blood drawing
-10 + 4 mm from normal
F. Measuring range
1 - 140 mm/h
G. Memory capacity
Up to 500 results
2) LIMITATIONS
A. Strongly lipemic or hemolytic samples may alter reading capability.
B. Sed rate values > 140 mm/h will be indicated with this mark only.
C. Temperatures outside the given range will be accepted as 15°C min and 32°C max