SRT 10/II
USER MANUAL
page 14 of 37
4.3 Display
A 40-character two-line LCD display with backlight, allows a continuous monitoring of the
analysis and a visualization of the results. On the display can be also shown a sample or a
system error message. Please note that the display can indicate more than one results for
each sample, according to dip-switch configuration set up by operator. To know which kind of
analysis results are referred to, please look in the bottom-left position on the display where
the message 30’, 1h or 2h will flash. On the display the operator can also see the symbols
shown below: These symbols suggest for each sample the analysis time remaining.
time remaining symbols
4.4 Reading Plate
One row of 10 test tube channels, numbered from 1 to 10.
4.5 Power Supply
The low voltage supply, built in accordance with international safety standards, offers a valid
solution to improving assistance service.
4.6 Printer
The printer is hooked up to the instrument with RS232 C outlet. Whenever an analysis is
completed, it prints out the ESR results following the loading sequence. The printer is
external so it can be easily replaced when necessary. It can also be replaced by other
printers with serial entry.
4.7 Test Tubes
The test tubes for this instrument are VACUETTE
®
(vacuum tube) specially made for the
SRT 10/II instrument. These tubes contain 3.2% sodium citrate and a vacuum which allows a
1.5 ml or
1.6 ml blood drawing.