![Greiner Bio-One Sed Rate Timer 10/II Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/greiner-bio-one/sed-rate-timer-10-ii/sed-rate-timer-10-ii_user-manual_2264278010.webp)
SRT 10/II
USER MANUAL
page 10 of 37
15 minutes analysis time
N°
Switch
Function
Position
1
Enable 15' analysis time.
ON
2
Enable 60' analysis time (works only if switch 1 = off)
x
3
Enable additional result formats (works only if switch 1 = off)
x
4
Enable temperature compensation
ON
5
Enable the printer
ON
6
Enable sedimentation curve graph
OFF
7
Enable the internal fan
ON
8
Enable power supply on printer connector (DPT100 printer)
ON
(where 'x' = don't care, has no effect)
30 minutes analysis time
N° Switch Function
Position
1
Enable 15' analysis time.
OFF
2
Enable 60' analysis time (works only if switch 1 = off)
OFF
3
Enable additional result formats (works only if switch 1 = off)
OFF
4
Enable temperature compensation
ON
5
Enable the printer
ON
6
Enable sedimentation curve graph
OFF
7
Enable the internal fan
ON
8
Enable power supply on printer connector (DPT100 printer)
ON