Accuracy
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows:
±
(a percentage of the r a fixed amount)
at 23
°
C
±
5
°
C (73.4
°
F
±
9
°
F), 0% to 75% relative humidity.
True RMS readings: Accuracies are specified from 5% to 100% of the range unless otherwise
specified. Frequency must be within the specified bandwidth for non-sinusoidal waveforms.
Crest factors are as follows:
• Crest factor < 3:1 at full scale
• Crest factor < 6:1 at half scale
AC Current
Range
Accuracy at
Accuracy at
Burden
50 Hz - 60 Hz
40 Hz - 1 kHz
Voltage
499.9
µ
A
±
(0.6% + 0.3
µ
A)
±
(0.8% + 0.4
µ
A)
0.15 mV/
µ
A
4999
µ
A
±
(0.6% + 3
µ
A)
±
(0.8% + 4
µ
A)
0.15 mV/
µ
A
49.99 mA
±
(0.6% + 0.03 mA)
±
(0.8% + 0.04 mA)
3.3 mV/mA
499.9 mA
±
(1.0% + 0.3 mA)
±
(1.0% + 0.4 mA)
3.3 mV/mA
4.999 A
±
(0.6% + 0.003 A)
±
(0.8% +0.004 A)
0.03 V/A
10.00 A*
±
(0.6% + 0.03 A)
±
(0.8% +0.04 A)
0.03 V/A
Range
Accuracy
Burden Voltage
499.9
µ
A
±
(0.2% + 0.4
µ
A)
0.15 mV/
µ
A
4999
µ
A
±
(0.2% + 4
µ
A)
0.15 mV/
µ
A
49.99 mA
±
(0.2% + 0.04 mA)
3.3 mV/mA
499.9 mA
±
(0.2% + 0.4 mA)
3.3 mV/mA
4.999 A
±
(0.2% + 0.004 A)
0.03 V/A
10.00 A*
±
(0.2% + 0.04 A)
0.03 V/A
* 10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off)
DC Current