Validation & Calibration
Page
55
of 82
PAH validation tolerance µg/l
Standard concentration
Lower tolerance
Upper tolerance
50
45
55
100
95
105
200
160
240
400
360
440
800
760
840
Table 7.2: Validation tolerance table
NOTICE
Immerse the sensor into the calibration as quickly as possible after
opening the ampoule to ensure the best performance and validity of the
calibration liquid.
Do not leave the sensor in the PAH calibration liquid more than 15
seconds since UV light degrades the PAH calibration liquid. After
approximately 15 seconds, the measurement value will be invalid due to
the degradation of the calibration liquid.
7.4.4 PAH validation procedure
Figure 7.11: G6111 PAH sensor & PAH calibration chamber