- 37 -
16. How to calibrate F-Mark2
The F-Mark2 needs to be calibration in order to read the crop mark correctly.
Please check the following items before performing the calibration of the F-Mark2.
The iMark Plus software is installed in the operation computer.
The F-Mark2 connected to the operation computer and CE7000-40.
The settings of CE7000-40 has been changed to the F-Mark2.
The angle of camera has been adjusted correctly.
The left side Push Roller of the CE7000-40 is in the correct position.
1. Lower the Media Set Lever of the CE7000-40, and loading the “Calibration Sheet”
into the CE7000-40.
When loading the calibration sheet into the CE7000-40,
load the calibration sheet so that the it is centered
between the left and right Push Rollers.
And change the position of right side Push Roller
according to the width of calibration sheet.
NOTICE
Never move the left side Push Roller position.