l
(
Recorder Output, Ion Gauge
Recorder Output, Thermocouple
Automatic ranging
Electrical Power Requirement
Customer specified preset at
factory via selector switch
Electrical Characteristics
Collector voltage to ground
Grid voltage to ground
Filament voltage to ground
Filament power available
Degas power available
Process control ran
�
e, ionization
gauge (each of 2
Process control ran e, thermocouple
gauge each of 2
Process control relay contact rating
Filament status relay contact rating
Thermocouple gauge mounting
Compatible Thermocouple Gauges
Physical:
Height
Width
Depth
Weight
Shipping weight
Mounting Hardware Supplied
I) 0 to -2.4V corresponding to front panel meter
indication of Oto 12 x I 0-x • Output impedance
is 2 Kohm.
2) 0 to -5V corresponding to IV /decade of pressure.
Output impedance is 2 Kohms.
0 to 2V corresponding to 0 to I Torr
(0
to 1.33
mbar) of air; output impedance is 1.86 ·Kohms
± 1%.
up
117% FS
down 9.5% FS
105 to 125 VAC, 50-60 Hz, I00W
210 to 250 VAC, 50-60 Hz, IO0W
90 to 110 VAC, 50-60 Hz, l 00W
ov
+ 180V nominal
+30V nominal
6 VAC, 6A
8 VAC, I0A(12A max)
-9
-4
3 x 10_9 Torr to 10 x 10 _4 Torr
3 x 10_7 mbar to 10 x 10 mbar
3 x l 0 Pa to l Ox 1Q-2pa
-3
I x IO _3 Torr to I Torr
I x 10_1 mbar to I mbar
I x IO . Pa to I 00 Pa
SPOT, 4A at I 15V AC, 2A at 230VAC, resistive
load
SPOT, 2A at I 15VAC, IA at 230VAC, resistive
load
I /8
11
NPT or .410 in. compression quick-connect
or suitable weld connection.
Hastings-Radist DV-6M, DV-6R, DV-36, and
DV-20
89 mm
(3.5
in.)
483 mm ( 19 in.) to fit standard 19 in. relay rack
292 mm ( I 1.5 in.) required for mounting
5.3 kg ( I 1.5 lb)
7 .5kg ( 16.5 lb)
Four I0-32x1/2 in. screws and four nylon washers
1-3