7.5.1 Temperature PT100/PT1000/Thermistor
The standard temperature sensors supplied (YS-EU, YS-CM etc) are all PT1000-
sensors. If you are connecting a standard sensor, select type
Temp.PT1000
.
If you are using a sensor purchased elsewhere, assign pins 1 to 4 and select the
appropriate sensor type in the software.
This connection can be configured in
LoggerSetup/SensorSlot
- SlotMode:
Temp.PT1000 [°C] or PT100 or Thermistor U
7.5.2 Sensors for temperature (Thermocouple)
The YL-T15 and YL-T16 allow connection of the thermocouple sensors. These can be
of any type from K, T, J, B, E, N, R or S. Connect the sensors into the logger with the
correct polarity.
This connection can be configured in
LoggerSetup/SensorSlot
- SlotMode:
Thermocouple [°C].
7.5.3 Combined Sensors for temperature / humidity
This connection can be configured in
LoggerSetup/SensorSlot
- SlotMode:
Temp/RH
[°C/%]
. Although humidity/temperature-sensors have are two separate sensors, they
only use one channel of the logger.
Humidity sensor life can be affected by environmental conditions. Read the Trouble
shooting section for more details.