7.6
Logger Setup – Configuring Alarms
Yoyo loggers have integrated alarm functionality. An LED will start to flash if thresholds
are exceeded.
Click on
Alarm
, to display the alarm settings.
Then click on the symbol of the
channel in order to change the
settings.
Select from the following Alarm Modes:
One alarm range:
Two alarm ranges:
Now set the values for
Threshold 1
and
Threshold 2
and define the
Hysteresis
Alarm
Alarm
No Alarm
Max
Min
Sensor
No Alarm
No Alarm
Alarm
Max
Min
Sensor
There is one alarm range for the selected
channel. If the sensor value is higher than
the (Alarm threshold1 ± Hysteresis) and
lower than (Alarm threashold2 ± Hysteresis)
the logger will start to alarm.
.
There are two alarm ranges for the selected
channel. If the sensor value is higher than
the (Alarm threshold1 ± Hysteresis) and
lower than (Alarm threashold2 ± Hysteresis)
there is no Alarm.
Otherwise the logger will start to alarm.
.