4. Operation
Recommendations during operation
When loading use even number of tubes arranged symmetrically (facing one another) to give the
unit even balance during operation. The opposite tubes must be filled up equally.
·
Centrifuge containers must not be filled over the capacity specified by the manufacturer.
·
Rotor must always be fixed securely. Stop the operation immediately with the
RUN/STOP
key if
any unusual noise occurs during acceleration which can be due to improper rotor fixation.
4.1. Check the power cord for any signs of damage and replace if necessary. Connect the power cord
to the mains outlet
. Set the power switch in
position I (ON) on the rear side.
4.2.
The centrifuge turns on and lid opens automatically.
The display shows the following readouts:
·
previously set time and speed in the upper line (
Set
);
·
mode indication (OPEN - lid opend, rotor stopped) and current speed - 0 RPM in the
lower line (
Actual
).
4.3.
Check the rotor and buckets for any signs of wear or corrosion and replace if necessary.
Insert EVEN number of tubes/microtest plates in rotor facing one another. The loading in
the opposite tubes must be equal.
which provides a safety earth (ground) terminal
Содержание LMC-3000 USA
Страница 1: ...Operating instructions Laboratory centrifuge LMC 3000 USA bio ...
Страница 2: ...Page 2 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...
Страница 14: ...Page 14 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...
Страница 15: ...Page 15 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...