Centrifuge

 

LMC

-3000

 USA

Operating Instructions

Version 

6

 - 

December 

20

11

4.4. 

Close the lid (the clicking sound of the lock) readings STOP and      (fig.3/5) in the lower 

line of the display indicate that the lid is closed.

Note! 

If any of these readings 

has not appeared on the display 

centrifugation cannot be started. Try to open and close the lid again.

4.5. 

With the “

p

” and “

q

” 

TIME

 buttons (fig. 3/1) set the required time interval (0-90 min, 

increment 1 min).

4.6. 

With the “

p

” and “

q

” 

RPM

 buttons (fig. 3/2) set the required speed (100-3000 RPM, 

increment 100 RPM). Note that the speed can also be adjusted during operation.

Note 

that some plastic tubes and microtest plates can be damaged at higher speeds. 

Refer to the tube material specifications to make sure that it will not get damaged at the 
set speed. Do not set speed higher than 2000 RPM when working with microtest plates!

4.7. 

Press 

RUN/STOP

 button (fig. 3/3) to start centrifugation. Blinking indication RUN and 

current speed is displayed in the lower line. The timer in the upper line starts countdown 
after the set speed is achieved (stable indication RUN). 

Note! 

If the rotor imbalance occurs causing vibration the centrifuge stops automatically 

(indication IMBALANCE). After the rotor is stopped open the lid and remedy the cause of 
imbalance. 

4.8. 

Centrifugation is stopped automatically after the set time elapses while display shows 

blinking indication STOP. A sound signal is emitted after full stop of the rotor (press 

RUN/STOP

 button (fig.3/3) to stop the signal). 

4.9. 

If necessary centrifugation can be stopped before the set time elapses by pressing 

RUN/STOP

 button. The set time interval will be shown on the display. 

4.10.

Press the button     

open 

(fig.3/4) and open the lid lifting it upwards with a hand (it is 

possbile to open the lid only when the rotor is stopped). Display shows OPEN.

4.11. 

At the end of operation set the power switch in position O (OFF) on the rear.

4.12.

Disconnect the power cord from the mains outlet.

Note! 

The electrical lid lock allows opening the lid only when the unit is connected to the 

mains and is turned on. Do not force the lid to open when the unit is switched off! 

Lid emergency opening

-

Disconnect the power cord from the mains outlet and allow the centrifugation to stop.

-

Slide the unit to the front of the bench to access the emergency opening slot on the underside of 
the unit (located in the front side). 

-

Avoid tilting the unit as this may cause spilling of the materials from the containers inside the 
unit. 

-

Insert a small screwdriver (or similar tool with diameter up to 3 mm) into the emergency opening 
slot in front of the dot on the label “Open” 

at a depth of 10-15 mm.

-

Move the lever from the left to the right hand side to release the lid lock.

 (STOP and      ) 

Содержание LMC-3000 USA

Страница 1: ...Operating instructions Laboratory centrifuge LMC 3000 USA bio ...

Страница 2: ...Page 2 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...

Страница 3: ...rifuge LMC 3000 USA Operating Instructions Contents Version 6 December 2011 1 Safety 4 2 General Information 6 3 Getting started 7 4 Operation 9 5 Specifications 11 6 Maintenance 13 7 Declaration of Conformity 15 ...

Страница 4: ...l ELECTRICALSAFETY Connect only to the mains with a voltage corresponding to that on the serial number label Ensure that the switch and plug are easily accessible during use Do not plug the unit into the main outlet without grounding and do not use extension lead without grounding Before moving the unit disconnect it from the mains To turn off the unit disconnect the power plug from the mains outl...

Страница 5: ...d except those recommended by the manufacturer check with the manufacturer that the proposed method will not damage the equipment Do not make modifications to the design of the unit BIOLOGICALSAFETY Without bioseal the centrifuge is not a biosafety system in accordance to EN61010 2 20 and cannot be used for centrifuging hazardous materials contaminated with toxic radioactive or pathogenic micro or...

Страница 6: ...l Information LMC 3000 USA is a modern benchtop laboratory centrifuge useful for sedimentation of cells bacteria yeast formed blood elements It provides operation with tubes and microtest plates LMC 3000 USA is designed for safe work metal protecting housing easy maintenance and wide application range in biochemical industrial and other type laboratories Version 6 December 2011 ...

Страница 7: ...Rotor R 2 on request Rotor support stand RR U on request Adapter sets BN 13 75 BN 13 100 and BN 16 100 for R 12 15 on request 3 3 Set up place the centrifuge on the even stable and clean surface plug the mains power cord into the socket on the rear and position the centrifuge so that there is easy access to the power switch and connector it is necessary to observe the safety area of 300 mm around ...

Страница 8: ... 3 4 Rotor replacement Hold the rotor with one hand and with the help of the supplied wrench 13 mm turn fixation nut fig 2 1 counter clockwise to release the rotor Replace the rotor and secure the new rotor carefully and turning the fixation nut tightly 1 Fig 2 Rotor replacement ...

Страница 9: ...se occurs during acceleration which can be due to improper rotor fixation 4 1 Check the power cord for any signs of damage and replace if necessary Connect the power cord to the mains outlet Set the power switch in position I ON on the rear side 4 2 The centrifuge turns on and lid opens automatically The display shows the following readouts previously set time and speed in the upper line Set mode ...

Страница 10: ... cause of imbalance 4 8 Centrifugation is stopped automatically after the set time elapses while display shows blinking indication STOP A sound signal is emitted after full stop of the rotor press RUN STOP button fig 3 3 to stop the signal 4 9 If necessary centrifugation can be stopped before the set time elapses by pressing RUN STOP button The set time interval will be shown on the display 4 10 P...

Страница 11: ... not more 30 sec Display LCD Chamber working diameter 335 mm Dimensions 495x410x235 mm Weight not more 13 5 kg Speed range 100 3000 RPM increment 100 RPM Operating voltage 120 V 50 60 Hz 120 W 1 A Operating conditions The unit is designed for operation in cold rooms incubators and closed laboratory rooms at ambient temperature from 4 C to 40 C and maximum relative humidity 80 for temperatures up t...

Страница 12: ...Page 12 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 Grant is committed to a continuous programme of improvement specifications may be changed without notice ...

Страница 13: ... parts inside the centrifuge chamber Rotors and other accessories are not autoclavable 6 4 Replacement of fuses fig 4 Disconnect from the outlet Remove the power plug from the rear of the unit Pull out the fuse holder by applying leverage in recess A Remove the fuse from the holder Check and replace with the correct fuse if necessary for 230V T1Aor for 120V T2A 6 5 Routine safety tests If routine ...

Страница 14: ...Page 14 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...

Страница 15: ...Page 15 Centrifuge LMC 3000 USA Operating Instructions Version 6 December 2011 ...

Страница 16: ...Grant Instruments Cambridge Ltd Shepreth Cambridgeshire SG8 6GB Tel 44 0 1763 260811 www grant co uk sales grant co uk Fax 44 0 1763 262410 Centrifuge LMC 3000 28713 6 ...

Отзывы: