ES-20
Operating instructions
Fuse replacement
Disconnect the unit from the mains. Disconnect the power cable from the socket on the rear
of the shaker-incubator. Open the fuse holder (fig.1/
z
). Replace with the correct fuse (for
230 V, – T1 A, for 120 V – T2 A).
6.3
Routine safety tests
If routine tests are to be made, we recommend a test of the integrity of the protective earth
conductor and an insulation test at 500 Vdc. Routine flash tests are not recommended for
any electrical equipment, because repeated high voltage tests degrade insulation materials.