8
Piece weight memory key.
Input piece weight using the 0 – 9 keys, then press this key to memorize
and press the keep where you want to save it, for exemple:
Press this key to select the PLU (M01 – M09) and press again to select the
PLU (M11 – M18)
Press this key to clear the piece weight window.
Press this key to display the total accumulation count, total accumulated
weight and total count.
To clear accumulation data, press and hold this key until buzzer beeps
twice.
STANDARD FUNCTIONS SETUP
To access to functions setting, press and hold the key, then press the
The weight window will show UF-1, press the key or key to select the
desired parameter (UF-1 to UF-8)
UF-1 Internal value / battery voltage
1. Press the key to enter.
2. Weight window will show the internal value (SPAN), piece weight window will show
the actual battery voltage and the total count window will display the the internal
value (ZERO).
2. Press the key to exit.
UF-2 Piece weight auto-average
1. Press the key to enter.
2. The display will show the message A A U G 1 (factory default).
3. Use 0 and 1 keys to enable or disable the auto-average function.
4. Press the key to confirm.
5. Press the key to exit.