11
ZP 5: 5 division not to display at zero
3. Press the to confirm
STANDARD CALIBRATION
1. Press and hold the key, then press the key.
2. Use the and keys to select ECF-1, ECF-2 or ECF-3
ECF-1 – Weight calibration
1. Press the to enter, display will show the internal value, press again the
key to calibration the zero weight after stable.
2. Use the 0 – 9 keys to enter the weight to be calibrated.
3. Place the required calibration weight on the platform.
4. Press the key to confirm.
5. The calibration procedure is completed.
6. Press the key to exit and return to previous setup mode.
ECF-2 – Zero calibration
1. Press the key to enter, display will show the internal value, press again the
Key to calibration the zero weight after stable.
2. Press the key to exit and return to previous setup mode.
ECF-3 – SPAN calibration
1. Press the key to enter.
2. Use 0 – 9 keys to enter the weight to be calibrated.
3. Place the required calibration weight on the platform.
4. Press the key to confirm.
5. The calibration procedure is completed.
6. Press the key to exit and return to previous setup mode.
Содержание INDICATOR SC2