1147-INS-LAB-RevD19  

12

  WARNING: Take into consideration the electromagnetic 

compatibility of the unit (e.g. disruptions to the power supply, 

radio frequency interference, etc.) Only use the unit indoors. 

To avoid inaccurate results due to electromagnetic interference 

between electrical and electronic equipment, do not use the unit 

near mobile phones or microwave ovens. Keep devices whose 

maximum power exceeds 2 W at least 11 feet (3.3 meters) from 

the blood pressure monitor.

  WARNING: Verify the accuracy of the blood pressure monitor 

every two years. Contact an authorized agent or laboratory for 

inspection / calibration service.

OPERATION PREPARATION

1.  Do not conduct any measurements if the temperature is low (below 

41°F / 5°C) or high (over 104°F / 40°C), or if the relative humidity 

is beyond the range of 15% to 90%, as this can lead to inaccurate 

readings.

2.  Take the measurement at room temperature in a quiet and stress-

free environment.

3.  Do not move yourself or the unit during the measurement. Do not 

speak during the measurement.

4.  Blood pressure varies naturally depending on the time of day and is 

affected by many factors. Blood pressure is usually highest at work 

and reaches its lowest level during sleep.

5.  Blood pressure measurements should be assessed by a physician 

or trained healthcare professional who is familiar with your medical 

history. If you use the unit and regularly record the results, keep your 

physician informed of any changes in your blood pressure.

6.  The performance of this device can be affected as severe 

arrhythmias such as atrial or ventricular premature beats or atrial 

fibrillation are presented during measurement.

7.  The blood pressure measurements conducted with this unit are 

equivalent to measurements obtained by a trained observer in 

accordance with the values achieved using the cuff/stethoscope 

auscultation method and are within the specified EN 1060-4 standard 

limits.

Содержание LUMISCOPE 1147

Страница 1: ...Blood Pressure Monitor Automatic Wrist Type Model 1147 Instruction Manual Read this manual before operating your Blood Pressure Monitor Save this manual for future use 1147 INS LAB RevD19...

Страница 2: ...LAY OVERVIEW 9 SETUP 10 BATTERY INSTALLATION 10 SETTING THE CLOCK 10 OPERATION 11 OPERATION SAFETY 11 OPERATION PREPARATION 12 MEASUREMENT POSTURE 13 APPLYING THE CUFF 13 TAKING A MEASUREMENT 14 MEMOR...

Страница 3: ...ides application recommendations or other useful information to ensure that you get the most from your product WARNING Important Read and understand these instructions before using the Blood Pressure...

Страница 4: ...with skin contact CONTRAINDICATIONS WARNING Do not use this appliance on infants children or persons who cannot express their own intentions WARNING Consult your physician before use during pregnancy...

Страница 5: ...e device or batteries with household waste Follow operating instructions Pumping Pumping failure Excessive body movement detected during measurement SYS and DIA readings detected are not reasonable Cu...

Страница 6: ...on Battery recycling symbol Restriction of Hazardous Substances REACH symbol Registration Evaluation Authorisation and Restriction of Chemicals REACH is a European Union regulation Power ON OFF Protec...

Страница 7: ...ts the highest pressure during heart contraction and diastolic pressure is the lowest pressure while heart is resting WHO ISH classification Your blood pressure measurement can be compared with classi...

Страница 8: ...w recommendations below 1 Keep taking and recording blood pressure measurement consistently at the same time every day to establish your blood pressure pattern 2 Wait at least thirty minutes after eat...

Страница 9: ...1147 INS LAB RevD19 9 DEVICE OVERVIEW Info This product includes one blood pressure monitor one cuff one instruction manual three AAA alkaline batteries and one storage bag case DISPLAY OVERVIEW...

Страница 10: ...r off mode press and hold the MEMORY button until the display shows a blinking year then press the POWER button for adjustment After current year is selected press the MEMORY button to set it and swit...

Страница 11: ...ed WARNING Avoid prolonged over inflation of the bladder to prevent physical injury WARNING If the cuff causes any discomfort during measurement press the POWER button to turn off the device immediate...

Страница 12: ...the measurement at room temperature in a quiet and stress free environment 3 Do not move yourself or the unit during the measurement Do not speak during the measurement 4 Blood pressure varies natura...

Страница 13: ...Sit with your hand relaxed your arm extended in front of you and palm facing up 7 Sit upright in a chair with your feet flat on the floor in a natural position and relax APPLYING THE CUFF 1 Ensure yo...

Страница 14: ...3rd bar from the bottom which falls within the yellow indication and grade 1 3 hypertension categories are indicated with 4 6 bars from the bottom which fall within the red indication 5 IPD Irregular...

Страница 15: ...button under memory mode until the display shows dEL and press the MEMORY button to show dEL ALL on the display then press the POWER button all readings in memory are erased after three beeps are hea...

Страница 16: ...emove any chemical residues remaining on the device 3 If necessary repeat steps 1 and 2 to ensure the device is clean 4 Wipe dry with a clean soft cloth to avoid water stains DISINFECTION WHEN CLEANIN...

Страница 17: ...asurement Refasten the cuff and measure again Insufficient pulse detected for measurement Refasten the cuff and measure again Detected SYS and DIA readings are not reasonable possibly as a result of t...

Страница 18: ...alkaline batteries Power Saving Auto off after 150 sec non operation Size 3 3 in x 2 9 in x 1 4 in 83 mm x 74 mm x 35 mm Weight 5 29 oz 150 g including batteries Cuff Range 5 31 in 8 66 in 13 5 cm 22...

Страница 19: ...ot limited to batteries accessories casters filters fuses gaskets lubricants and charts 4 Accessories or parts not provided by GF 5 Charges by anyone for adjustments repairs replacement parts installa...

Страница 20: ...mation can be found on our website www grahamfield com 2009 GF Health Products Inc All Rights Reserved Lumiscope is a trademark of GF Health Products Inc One Graham Field Way Atlanta GA 30340 3140 GF...

Отзывы: