1031FP-INS-LAB-RevA11
16
INSTALLATION OF 1010S005 HYDRAULIC PUMP
Removal of Existing Pump
1. Follow the instructions in this
section to remove the existing
pump, shown at near right. If
installing pump for the irst time,
proceed to
INSTALLATION OF
NEW PUMP
.
2. Remove the top of the exist-
ing pump from the upper boom
bracket, as shown at mid-right:
Use two
9/16
"
wrenches to remove
the hardware that secures top
of pump to upper boom bracket.
This will free the pump to rotate
downward.
3. Remove the bottom of the exist-
ing pump from the lower mast
bracket, as shown at far right. Use
two
9/16
"
wrenches to remove the
hardware that secures bottom of
pump to lower mast bracket. This
will free the pump — take care
that it does not fall.
lower mast
bracket
upper boom
bracket
boom
mast
existing pump
existing pump
detach existing pump
from upper boom
detach existing pump
from lower mast
Содержание Lumex LF1030