JB02020-INS-LAB-RevA20 • Fingertip Oximeter User Manual 

12

5  OPERATION INSTRUCTIONS

Before use: surface-clean sensor with a soft cloth damped with 

a solution of 75% isopropyl alcohol, then surface-clean with a 

cloth damped ONLY with clean water and dry with a clean, soft 

cloth.

Open Oximeter and Position Finger

1.  Open clip, place finger between the rubber cushions as 

shown above, and gently close clip on finger.

2.  After two seconds, the Oximeter will automatically power on 

and begin to measure.

3.  The display screen is shown above; short press the display 

key to shift the display direction; long press the display key to 

see the measured PI value.

Info: During measurement, if SpO

2

 or Pulse Rate value 

exceeds the preset limit (see Preset Over-limits in TECHNICAL 

SPECIFICATIONS section), that measured value which exceeds 

limit will flash on the screen.

Содержание John Bunn JB02020

Страница 1: ...ximeter User Manual Important Do not operate the Finger Pulse Oximeter without first reading and understanding this manual Save this manual for future use Info The most current version of this manual...

Страница 2: ...8 CONTENTS PACKING LIST 8 JB02020 FINGER PULSE OXIMETER FEATURES 9 DECLARATION OF CONFORMITY 10 4 SETUP BATTERY INSTALLATION 11 5 OPERATION INSTRUCTIONS 12 CORRECT MEASUREMENT PROCEDURES 13 6 SERVICE...

Страница 3: ...tended to measure the pulse rate and functional oxygen saturation SpO2 through a patient s finger It is applicable for spot checking SpO2 and pulse rate of adult and pediatric patients in homes and me...

Страница 4: ...e the same as or similar to the following or specific to the operations WARNING Indicates a potential hazard situation or unsafe practice that if not avoided could result in death or serious injury CA...

Страница 5: ...o visible damage that may affect patient or user s safety or measurement performance If there is visible damage do not use the device Thereafter inspect the device at least once a week WARNING DO NOT...

Страница 6: ...of the oximeter For this reason ensure that all external devices operated in the vicinity of the oximeter comply with the relevant EMC requirements Mobile phones X ray equipment and or MRI devices ar...

Страница 7: ...it immediately allow at least fifteen minutes for the Oximeter to reach ambient temperature s NOTICE DO NOT operate the button on the front panel with sharp materials or sharp point s NOTICE DO NOT u...

Страница 8: ...er User Manual 8 3 PRODUCT DESCRIPTION AND FEATURES display screen display key rubber cushion JB02020 Finger Pulse Oximeter Contents Packing List 1 One JB02020 Fingertip Oximeter 2 One Lanyard 3 Two A...

Страница 9: ...INS LAB RevA20 Fingertip Oximeter User Manual 9 JB02020 Finger Pulse Oximeter Features S L M Display type Displays SpO2 PR PI Auto on off PR and PI shifts Pulse bar Over limits indicator Two direction...

Страница 10: ...electrical equipment Part 1 General requirements for basic safety and essential performance BS EN ISO 9919 2009 or the equivalent ISO 80601 2 61 2011 Medical electrical equipment Part 2 61 Particular...

Страница 11: ...eries into the battery compartment as shown above Note polarity markings and ensure batteries are correctly installed or device may not function 3 Replace the cover s NOTICE Change the batteries when...

Страница 12: ...ace finger between the rubber cushions as shown above and gently close clip on finger 2 After two seconds the Oximeter will automatically power on and begin to measure 3 The display screen is shown ab...

Страница 13: ...sh or paint may affect the measuring accuracy and too long a fingernail on measurement finger may cause failure of measurement or inaccurate result Vigorous exercise and electrosurgical device interfe...

Страница 14: ...l only or device failure and health hazard could occur Maintenance s NOTICE In order to ensure a long service life adhere to the following maintenance instructions Change batteries when low voltage in...

Страница 15: ...e device When the oximeter is taken from a cold environment to a warm and humid environment do not use it immediately allow at least fifteen minutes for the Oximeter to reach ambient temperature Dispo...

Страница 16: ...ly as described in OPERATION INSTRUCTIONS try again Unable to turn on device no display 1 Patient must relax as described in OPERATION INSTRUCTIONS try again 2 Change batteries Display direction doesn...

Страница 17: ...nm Infrared light 890 nm Maximal average optical output power 2mW SpO2 display range 35 100 SpO2 measuring accuracy 2 for SpO2 range from 70 to 100 Pulse Rate measurement PR display range 30bpm 240bp...

Страница 18: ...perating current 40mA Environmental Conditions Temperature Operating 41 F 104 F 5 C 40 C Storage 4 F 140 F 20 C 60 C Humidity Operating 30 80 Storage 10 95 Atmospheric pressure Operating 70kPa 106kPa...

Страница 19: ...nally powered equipment Degree of protection against electric shock Type BF applied parts Degree of protection against harmful solid foreign objects and ingress of liquid IP22 with protection against...

Страница 20: ...ion Index Pulse Strength Bar Graph Low Battery Voltage CE Mark Serial Number Date of Manufacture Authorized Representative in the European Community Dispose of this Device in Accordance with Local Law...

Страница 21: ...the applicable warranty period and which upon examination by GF or its authorized representative prove to be a warranty item This limited warranty is not transferable Within the guidelines set forth i...

Страница 22: ...parts installation or other work performed upon or in connection with such products which are not expressly authorized in writing in advance by GF 6 Any labor or shipping charges incurred in the repl...

Страница 23: ...ed on any representation assertion guarantee warranty collateral contract or other assurance except those set out in this document Some states do not allow the exclusion of certain remedies in those i...

Страница 24: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 24 APPENDIX I EMC This equipment meets the requirements of IEC 60601 1 2 2014 Table 1...

Страница 25: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 25 Table 1 continued...

Страница 26: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 26 Table 2 continued...

Страница 27: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 27 Table 2 continued...

Страница 28: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 28 Table 2 continued...

Страница 29: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 29 Table 3 continued...

Страница 30: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 30 Table 3 continued...

Страница 31: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 31 Table 3 continued...

Страница 32: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 32 Table 3 continued...

Страница 33: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 33 Table 3 continued...

Страница 34: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 34 Table 4 continued...

Страница 35: ...JB02020 INS LAB RevA20 Fingertip Oximeter User Manual 35 Table 4 continued...

Страница 36: ...tion can be found on our website www grahamfield com 2020 GF Health Products Inc All Rights Reserved John Bunn is a trademark of GF Health Products Inc One Graham Field Way Atlanta GA 30340 3140 GF He...