VIC429-INS-LAB-RevA21 22
4 PREVENTIVE MAINTENANCE FOR THE USER
Component
Cleaning Procedure
Schedule
Cleaning Agent *
Special Notes
Pads / Mattresses
Wipe with damp cloth to
remove any foreign material
After
each
use
Routine hospital grade
disinfectants, soap and
water
Use only medium
strength cleaners
Do not steam clean
or pressure wash
Chair
Wipe with damp cloth to
remove any foreign material
Routine hospital grade
disinfectants, soap and
water
Lubricate pivot
points after
cleaning
Mechanical chair
components
Wipe with damp cloth to
remove any foreign material
Routine hospital grade
disinfectants, soap and
water
Lubricate pivot
points after
cleaning
Mechanical
accessories
Wipe with damp cloth to
remove any foreign material
Routine hospital grade
disinfectants, soap and
water
Lubricate pivot
points after
cleaning
Procedure
Schedule
Material
Lubricate all actuator mounting pins
Every 6
months
Lithium-based grease
NEVER LUBRICATE COLUMN
Inspect all fasteners to ensure proper fit, position and
tightness, including nuts, bolts, etc.
Every 3
months
Proper size wrench and screwdriver
Inspect all surfaces and remove any sharp or burred
areas; apply touch-up paint where required
Metal file, proper color paint (specify color
when ordering)
* Disinfecting and Cleaning Upholstery - ALWAYS follow manufacturer’s recommended dilution
Disinfectants for
vinyl products
Phenolic disinfectants are the best choice for vinyl
Properly diluted quaternaries are also acceptable for vinyl
Quaternary / Isopropyl disinfectants ARE NOT recommended for vinyl
Disinfectants
for urethane
products (Standard
Upholstery)
Quaternary disinfectants are recommended for urethane
Quaternary / Isopropyl disinfectants are recommended for urethane
Phenolics SHOULD BE AVOIDED on urethane
Disinfectants for all
products
All fabrics may be cleaned with a 1:10 dilution of household bleaches containing 5.25%
sodium hypochlorite as recommended by the Centers for Disease Control in Atlanta,
Georgia; there is no harmful effect on the fabric
Disinfectants applied at full concentration or in highly concentrated solutions will decrease
the useful life of fabric
Iodophor-type disinfectants used on fabric may result in staining
Soils or Stains
Use neutral soapsuds and lukewarm water; DO NOT use harsh cleansers, solvents or
detergents
Hard-To-Clean
Spots
Use standard household / vinyl cleansers and a soft bristle brush on troublesome spots or
stains; presoak heavy, dried-on soil
Laundering
Laundering Vinyl-laminated, Polyurethane-coated, or Rubber-coated fabric IS NOT
recommended; laundering may substantially decrease the useful life of the fabric
s
NOTICE — POSSIBLE EQUIPMENT DAMAGE HAZARD: Steam cleaning and pressure
washing of chair is not recommended and can void warranty.
Info: For more detailed information, contact GF Health Products, Inc. at 1.770.368.4700.
Info: GF Health Products, Inc. offers customized Preventative Maintenance Service Programs
for Hausted products; contact your GF sales representative for further information.