GPI TM050 Скачать руководство пользователя страница 3

3

Conditioned Signal Output 

Module Wiring

This conditioned signal output module can be 

wired to provide an open collector signal output 

or 6-volt square wave output. 

Open Collector Signal Output

To achieve an open collector signal output, 

reference Wiring Diagram 1. The terminal block 

is located on the back side of the module. The 

module is factory assembled for open collec-

tor signal output. Please provide the (820 ohm 

minimum) resistor. 
Ten feet (3 m) of cable is provided with the 

module. Trim it to desired length or extend it as 

necessary. Distances up to 5,000 feet (1,524 m) 

can be achieved for open collector signal output.

Square Wave Output

To achieve square wave output, reference Wir-

ing Diagram 2 and use an Electronic Digital 

Meter Battery Kit (sold separately) for battery 

power. The terminal block and battery location 

are located on the back side of the module. 

Access as follows:

1.  Remove the four Phillips-head screws from 

the front of the module and lift the module 

from the turbine.

2.  To change terminal block connections, 

loosen the appropriate screws. Reconnect 

the wires in the proper positions and tighten 

the screws.

3.  Install the batteries. Make sure the positive 

post is in the correct position.

4.  Position the module on the turbine housing. 

To avoid moisture damage, make sure the 

seal is fully seated. Tighten the four screws 

on the front of the module.

Ten feet (3 m) of cable is provided with the 

module. Trim the cable to desired length or 

extend it as necessary. 

Verify Meter Accuracy

Before using, check the meter’s accuracy and 

verify calibration.
1.  Make sure there is no air in the system  by 

starting the flow until it runs steadily. Then, 

stop or divert the flow using a valve or nozzle.

SQUARE WAVE OUTPUT

Wiring Diagram 2

3

2

1

4

White

Black

Common
(Ground)

Terminal

Block

Customer

Interfacing

Equipment

6 volt Square

Wave Signal

The terminal block is identified as follows:

Pin #1 6 volt square wave

Pin #2 +9 to 35 volt DC Input (not used)

Pin #3 Common Ground

Pin #4 Open Collector signal Output (not used)

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание TM050

Страница 1: ...inches of some electric motors or other sources of electronic noise INSTALLATION Connections Install your meter in line either horizontally or vertically or at the end of the hose adjacent to the nozz...

Страница 2: ...curls can adversely affect accu racy Correct the problem prior to turbine installation NOTE Do not over tighten the flange bolts This may cause the gasket to be compressed into the flow stream and ma...

Страница 3: ...ule and lift the module from the turbine 2 To change terminal block connections loosen the appropriate screws Reconnect the wires in the proper positions andtighten the screws 3 Install the batteries...

Страница 4: ...eplaceable batteries When display becomes dim faded or the low battery message appears see below the batteries need to be replaced Reference the Maintenance Section for details Factory and Field Calib...

Страница 5: ...ct When satisfied with the value press both CALIBRATE and DIS PLAY buttons simultaneously CALEnd will be displayed and unit will go back to normal operation less the FAC factory calibration icon 4 All...

Страница 6: ...becomes dim blank or the low battery message appears see below replace the batteries as follows 1 Remove the four Phillips head screws from the face of the meter and lift the faceplate from the turbin...

Страница 7: ...ch 222 kg 381 kg 381 kg 1 inch 299 kg 626 kg 666 kg 2 inch 354 kg 807 kg 875 kg Dimensions cm W x H x L Without Fitting With NPT Fitting 5 3 x 6 4 x 10 9 5 3 x 6 9 x 15 2 5 3 x 6 9 x 11 2 5 3 x 7 3 x...

Отзывы: