b�  If the meter registered less than 

the quantity in the container, turn 
the calibration screw clockwise� 
If the meter registered more 
t h a n   t h e   a m o u n t   i n   t h e 
container, turn the calibration 
screw counterclockwise�

5�  Empty the calibration container 

and repeat steps 2 to 4 until the 
meter registers the quantity in the 
container�

6�  Install the nylon washer and calibra-

tion screw cover�

The meter’s strainer should be cleaned 
at regular intervals, especially when low 

flow is detected.

Clean or Replace Strainer

1�  Remove the nuts and bolts at the 

inlet fitting� Remove the fitting, 
O-ring, and strainer�

2� 

Using  a  fine  brush,  clean  the 

strainer� Replace the strainer as 
necessary�

3�  Wipe clean the inlet, housing, and 

O-ring groove� Coat the O-ring with 
oil or light grease� Make sure the 
O-ring is fully seated� Replace the 
strainer�

4� 

Position the inlet fitting in the desired 

orientation and tighten the nuts and 
bolts to 60 - 72 in-lb�

Figure 4

The large meter display represents the 
Batch Total for each fuel delivery� Before 
dispensing, reset the Batch Total to zero 

by turning the knob. (Figure 3)

The small display represents the 
Cumulative Total of all fuel deliveries 
and cannot be reset� 

CALIBRATION

The meter is accurately calibrated at 
the factory for use with kerosene grade 

(JET A)

� Due to differences in viscosity 

and flowrates, the meter may require 

recalibration to measure other fuels or 
to adjust for inaccuracies�
1�  Purge air from the meter and fuel 

system by dispensing fuel into a 

container  until  a  full  flow  occurs. 

Close the nozzle�

2�  Reset the meter counter to zero by 

turning the reset knob. (Figure 3)

3�  Pump into a graduated calibration 

container  to  a  specified  quantity. 

For the greatest accuracy, be sure 
the container is placed on a level 

surface and a consistent flow rate is 

used� When topping off the calibra-
tion container, use a quick-open and 
quick-close method until the mark 
is reached�

4�  Compare the meter display to the 

quantity in the container� If the 
meter display does not register the 
quantity on the container, adjust the 
meter by performing the following:

a�  Gain access to the recessed 

calibration screw by turning the 
calibration screw cover and seal 
counterclockwise� Remove the 

cover and nylon washer. (Figure 4)

MAINTENANCE

Figure 3

Содержание MR 5-30

Страница 1: ...s on receiving your GPI MR5 30Meter Wearepleasedtoprovide you with a meter designed to give you maximum reliability and efficiency Ourbusinessisthedesignmanufacture and marketing of liquid handling agricultural andrecreationalproducts We succeedbecauseweprovidecustomers withinnovative reliable safe timely and competitively priced products We pride ourselvesinconductingourbusinesswith integrity and...

Страница 2: ... manual DA N G E R DANGER indicates a hazardous situation which if not avoided will result in death or serious injury WARNING WARNING indicates a hazardous situation which if not avoided could result in death or serious injury CAUTION CAUTION indicates a hazardous situation which if not avoided may result in minor or moderate injury It is your responsibility to know and follow applicable na tional...

Страница 3: ... the two screws that hold the faceplate in place and remove faceplate Figure 1 3 Remove two screws to release the counter assembly Figure 2 4 Remove the bevel gear from the center of the coverplate Remove the eight pairs of nuts and bolts that hold the coverplate assembly in place 5 Now you can rotate the coverplate to the desired orientation Reassemble in reverse order making sure the O ring is f...

Страница 4: ... reset the Batch Total to zero by turning the knob Figure 3 The small display represents the Cumulative Total of all fuel deliveries and cannot be reset CALIBRATION The meter is accurately calibrated at the factory for use with kerosene grade JET A Due to differences in viscosity and flowrates the meter may require recalibration to measure other fuels or to adjust for inaccuracies 1 Purge air from...

Страница 5: ...ween coverplate Remove coverplate and inspect and housing for damaged missing or incorrectly seated seal Replace as required 3 Leakage at fittings Remove fittings and inspect for damaged missing or incorrectly seatedseals Replaceasrequired 4 Leakage at threads Remove meter and reseal all threaded connections with thread tape or pipe thread sealing compound approved for use with flammable liquids S...

Страница 6: ... and hardware 126530 02 Gallon Faceplate Kit includes faceplate decal and hardware 126530 03 Litre Faceplate Kit includes faceplate decal and hardware 126530 01 Bezel Kit includes hardware 126530 07 Knob Assembly Kit 126530 12 Fitting Kit for 3 4 inch NPT 2 Fittings 2 Fitting O Rings 1 strainer and hardware 126530 13 Fitting Kit for 3 4 inch BSPP 2 Fittings 2 Fitting O Rings 1 strainer and hardwar...

Страница 7: ...andling requirements you will need to follow covering the transportation and handling of fuel transfer equipment Before packing for shipment make sure the meter is thoroughly drained and free of fuel and vapors Kits and Accessories continued Kit No Part No Description Do not return meters or parts without specific authority from the GPI Customer Service Department Due to strict regulations governi...

Страница 8: ...f Manufacturer The warranty shall extend to the purchaser of this product and to any person to whom such product is transferred during the warranty period This warranty shall not apply if A the product has been altered or modified outside the warrantor s duly appointed representative B the product has been subjected to neglect misuse abuse or damage or has been installed or operated other than in ...

Отзывы: