5. CALIBRATION

CALI 5

Note:

  Provides calibration adjustment up to +/- 5% of

reading. You will need to calculate the percentage
adjustment you require. If the meter is reading over you will
need to enter a minus figure. If the meter is reading under you
will need to enter a plus figure.

51 CALIBRATION = CALI 51 ( adjust calib/- 5%)

b) Press the 

DISP

 button 3 times to reach “CALI 5”

c) Press 

RESET

 “CALI 51” will appear.

d) Press and hold down 

RESET

 for 2 seconds ( program will

    appear and the first digit will flash)

e) Press 

RESET

 to change number.

   Press 

DISP

 to change the decimal point position, or go

    to the next number.
    Press the RESET button to set the negative value.

f) Press and hold down 

RESET

 until “program” disappears

    to accept setting.

g) Press 

RESET

 for five seconds to return the meter to the

    operating mode.

6. CONFIGURATION

CONF 6

6 CONFIGURATION = CONF 6 (configuration details)

b) Press the 

DISP

 button 4 times to reach “CONF 6”

c) Press 

RESET

 “MOD” plus model number  will appear.

d) Press 

RESET

 “REL” plus hardwear release number will

    appear.

e) Press 

RESET

 the screen will flash between “SOFT” and

    the softwear release no’s (at 2 second intervals)

f) Press 

RESET

 the screen will flash between “SEr” and the 6

   figure serial number (at 2 second intervals)

g) Press 

RESET

 “MAN” plus manufacturer number

h) Press 

RESET

 the screen will flash between “CONFIG”

    and the configuration number (at 2 second intervals).

i) Press 

RESET

 for five seconds to return the meter to the

   operating mode.

7. DIAGNOSTICS

dIA 7

a) Press and hold down 

RESET

 for 5 seconds until  “MEtr 1”

    appears.

71 DIAGNOSTICS = dIA 7 (diagnostic mode)

b) Press the 

DISP

  button 5 times to reach “dIA 7”

c) Press 

RESET

 “ dIA 71”  will appear.

d) Press and hold down 

RESET

 for 2 seconds ( Program will

     appear then all the segments should be displayed).

e) Press the 

DISP

 button “diSP” ( to check that the display

    button is working properly).

f) Press the 

RESET

  button “RESEt” ( to check that the reset

   button is  working properly).

g) 

Press and hold down 

RESET

 for 2 seconds ( Program

    will disappear (check the condition of the reed switch).

The condition of the reed switch can only be checked
when there is liquid flowing through the meter. The reed
switch should be reading between 50 - 80.

h) Press 

RESET

 for five seconds to return the meter to the

    operating mode.

d) Press 

RESET

 to select default time (5-20 seconds)

e) Press and hold down 

RESET

 until “program” disappears

    to accept setting.

f) Press 

RESET

 for five seconds to return the meter to the

    operating mode.

Step 1. Reset the batch reading to zero.

Step 2. Dispence at least 5 liters, using a constant flow rate,
            into a graduated calibrated container.

Step 3. Take the reading from the IM50.

Step 4. Calculate the % error as follows.

 

% error = (( container volume - IM50 volume) / container volume) * 100

Step 5. Enter the % error into the meter as calculated, if the
            value is negative enter a negative number

a) Press and  hold down 

RESET

 for 5 seconds until  “totL2”

    appears.

a) Press and  hold down 

RESET

 for 5 seconds until  “totL2”

    appears.

c) Press and hold down 

RESET

 for 2 seconds ( program will

    appear)

b) Press the 

DISP

 button 2 times to reach “tiME 43”

a) Press and  hold down 

RESET

 for 5 seconds until  “totL2”

    appears.

43 TIME = tiME 43 (time to revert to default display)

Note: The display shows ( Hxx   rYY ) when fluid flows through
          the meter. YY is the sensitivity of the reed switch.

Содержание LM50DB

Страница 1: ...er has incorporatedtheovalrotorprincipalintoitsdesign Thishas proven to be a highly reliable and highly accurate method of measuring flow providing exceptional repeatability and high accuracy GPI reco...

Страница 2: ...nic computer module is non repairable The meter will need to be replaced if found to be damaged or faulty 1 Remove the 8 x philips head screws 1 from the meter cover 2 Remove the meter cover 2 and o r...

Страница 3: ...oint 25 change non resetable accumulative total units 26 change non reset accumulative total decimal point 22 change batch total decimal point 27 change K factor 31 change flowrate units 32 change flo...

Страница 4: ...1 1 decimal place 27 K FACTOR kFac 27 normal setting is 112 00 pulses b Press RESET 7 times kFac 27 will appear c Press and hold down RESET for 2 seconds program will appear To change the K factor pr...

Страница 5: ...nds program will appear e Press RESET to select sleep setting no sleep sleep 5 f Press and hold down RESET until program disappears to accept setting f Press and hold down RESET until program disappea...

Страница 6: ...1 DIAGNOSTICS dIA 7 diagnostic mode b Press the DISP button 5 times to reach dIA 7 c Press RESET dIA 71 will appear d Press and hold down RESET for 2 seconds Program will appear then all the segments...

Страница 7: ...7 NOTES...

Страница 8: ...n d to th e p u r c h a s e r o f th is p r o d u c t a n d to a n y p e r s o n to w h o m s u c h p r o d u c t is tr a n s fe r r e d d u r in g th e w a r r a n ty p e r io d T h e w a r r a n ty...

Отзывы: