11. Troubleshooting
No.
Problem
Reason
Measure
4
Air leakage
The air bag in the sleeve is damaged
Purchase a new sleeve
The air line to the air bag is disconnected
Reconnect the air line
5
Water leakage
The water bag in the sleeve is damaged
Purchase a new sleeve
The water line to the water bag is disconnected
Reconnect the water line
Male/Female check valve malfunction
Replace the connector
The water pump is disconnected from the water
lines
Power off the device, remove all water from the
sleeve. Use an absorbent cloth and connect the
water lines to the water pump. Make sure the
connections are secure and that there are no leaks.
6
LCD screen not showing
images
Low battery
Charge the battery
Screen damaged
Contact manufacturer
Cable not connected
Connect the cable
When the unit begins to beep, the air pump duration is too long. Press the On/Off button to turn off.
Press the button after three seconds to restart the unit.
Caution:
Page 27-1