11. Troubleshooting
No.
Problem
Reason
Measure
1
The wrap is not deflating
properly
The air line is twisted or bent, blocking the air from
releasing.
Straighten the air line
Air bladder blocked
Unplug the air line connector and shake the sleeve
to release the air from air bladder
Normal open solenoid malfunctioned
Power off the device, contact manufacturer
2
The wrap is not inflating
properly
The battery power is low
Immediately power off the device and recharge the
battery
Air connector not connected
Connect air connector
Air connector O ring damaged
Replace the connector
The air line is twisted or bent
Straighten the air line
Air bladder leaking
Change a new wrap
Normal open or Normal close solenoid
malfunctioned
Power off the device, contact manufacturer
Air pump malfunctioned
Power off the device, contact manufacturer
3
The device is not pumping
water
The suction tube and water return tube have
disconnected
Immediately power off the device and reconnect
them
The suction tube blocked by ice
Unplug the suction tube and remove obstacle from
tube
Water filter too full
Unplug the water line and clean out the filter with
running water
The water lines have disconnected
Immediately power off the device and find where
the lines have disconnected and reconnect them
Malfunctioned water pump
Contact manufacturer
Check valve malfunction
Replace the connector
Page 27
Содержание 700-2
Страница 12: ...1 2 3 4 5 6 Chamber 4 5 6 Chamber 1 2 3 Distal Proximal Full limb Chamber 1 2 3 4 5 6 1 2 3 4 5 6 Page 8 1...
Страница 13: ...6 Control panel description 1 2 4 3 5 6 Full limb Page 8 2 1 Air bladder effective area setup...
Страница 33: ...6 Fold top of the reservoir 3 times around the stiffening blade 7 Close the buckle 7 Operation Page 20 1...
Страница 41: ...Extended splitter can be used to attach up to four sleeves Extension 7 Operation Page 22 1...