GoyaLab SAS
www.goyalab.com
Institut d’optique d’Aquitaine, rue François Mitterrand
- 33400 Talence
–
France
Version June 20
g.
Change of Device
The app can be used with all the product offered by GoyaLab. You can
switch from one device to another by pressing name of the device.
Using the IndiGo with the UV MODULES
Precaution for use
As a standard mode, the UV module is equipped with UV A and/or UV B LED that can be dangerous to
eyes and skin when directly exposed to this type of radiation. Therefore, please wear appropriate
safety glasses, cover your skin and avoid direct exposure to your skin and eyes when the UV LEDs are
emitting.
Installation
Before installing the UV module, switch off the IndiGo to
avoid electrical damage and connection issues.
Then place the 10 metal pins in front of the 10 holes on
the front face of the IndiGo. Use two M3x10 screws
supplied with the module to fix the UV module.
You can now switch on the IndiGo and take
measurements.
How it works
a.
UV A version
In the standard UVA version, the UV module can be provided with one of the
following wavelengths: 365nm, 375nm, 385nm or 395nm.
The left button (1 on the opposite figure) will activate LEDs
a)
and
d)
in high
power mode with a power greater than 350mW per LEDs.
The right button (2 on the opposite figure) will activate the LEDs
b)
,
c)
,
e)
and
f)
in low power mode
with 20mW per LED.