Model 75-800 & 75-800-AR Noble Gas Purifier
Rev. 7 05/19
11
8.2 Operation
Operate the system at 400 °C. Once every week, turn the temperature up to
450 °C for one (1) hour. After one hour, reduce the temperature to 400 °C and
return to running samples. This procedure will eliminate any H
2
that may have
accumulated during that week.
9
Waste Disposal
This procedure is for GOW-MAC Mini Purifier Part No. 180-542 (Helium) and
180-542-AR (Argon).
If the purifier is used, then the getter material inside the purifier is already in a non-
reactive state, (i.e. will not react with air).
If the material is still in a reactive state (i.e. the purifier is not completely "used up"),
it still will not sustain combustion in air except at temperatures over 350 °C,
and with a continued access to air.
The operator should place a VCR cap over the inlet/outlet fittings for sufficient
precaution before disposal. It is then acceptable to deposit in an approved landfill,
in accordance with local, state and/or federal regulations. When disposed of in this
manner, the waste poses no known environmental problems.
10
Replacement Parts
Part No.
Description
121-100
Fuse, 1.6 A, 250 V (230 V)
121-160
Fuse, 3 A, 250 V (115 V)
122-114
Relay, solid state
124-175
Platinum probe, RTD 100
Ω
124-194
Heater, 110 V, 100 W
124-195
Heater, 240 V, 150 W
124-261
Temperature Controller, Watlow EZ-Zone
126-184
Terminal Block, 5-position
127-407
Cord, 7' 6", detachable
129-345
Receptacle with power switch and fuse holder
141-452 Feet
180-542
Mini-purifier with 1/4" VCR fittings (for all models except
75-800-AR & 75-802-AR)
180-542-AR Mini-purifier with 1/4" VCR fittings (only for models 75-800-AR &
75-802-AR)