Model 75-800 & 75-800-AR Noble Gas Purifier

 

Rev. 7  05/19

9

 

6.2  Responding to a Displayed Message/Error Code

 

 

  An active message will cause the display to toggle between the normal 

settings and the active message in the upper display and [

Attn

] in the lower 

display.

 

 

  Your response will depend on the message and the controller settings. Some 

messages, such as Ramping and Tuning, indicate that a process is underway. 

If the message was generated by a latched alarm or limit condition, the 

message can be cleared when the condition no longer exists. If an alarm has 

silencing enabled, it can be silenced.

 

 

  Push the Advance Key to display [

.9nr

] in the upper display and the message 

source (such as [

L.hi

]) in the lower display.

 

 

  Use the Up and Down keys to scroll through possible responses, such as 

Clear 

[

CLr

] or Silence [

S.L

]. Then push the Advance or Infinity key to execute the 

action.

    

Er. 1

  -  

Error Input 1:

 The sensor input generated a value lower than the 

allowable signal range. Enter a valid input.

    

100

 

 -  

Device Error:

 Controller displays internal malfunction message at 

Power Up.

Set Point Settings

 

 A.  To set the set point: turn the unit “on” and then push the Up or Down arrow to 

  adjust to the desired operating temperature.

Settings

 

8.1  Initial Start-Up

 

    A.  The purifier should be purged with an inert gas (minimum purity 99.99%)

 

before activating the material (He, Ne, Ar, Kr, Xe).

 

    B.  Start a flow through the unit for at least 10 minutes. Use a minimum of 

10% - 20% of specified flow rate for the unit. Use a continuous purge flow 

during activation.

CAUTION

THE PURIFIER MATERIAL SHOULD NEVER BE 

HEATED ABOVE 150°C IN AIR.

Содержание 75-800

Страница 1: ... 800 AR Noble Gas Purifier 75 800 120 V 50 60 Hz 75 802 230 V 50 60 Hz 75 800 AR 120 V 50 60 Hz 75 802 AR 230 V 50 60 Hz May 2019 Rev 7 READ INSTRUCTIONS BEFORE OPERATING 277 Brodhead Rd Bethlehem PA 18017 8600 U S A Tel 610 954 9000 ...

Страница 2: ...GOW MAC Instrument Co 2 ...

Страница 3: ...grade the packing When not in use CAP both the inlet and outlet fittings 1 4 VCR at the cabinet panel If minimal exposure to air at 150 F 66 C has occurred one may try to recondition the purifier Follow the instructions in this manual or contact GOW MAC at 610 954 9000 for assistance Models Affected 75 800 75 800 1 75 800 2 75 800 ADD 75 800 AR 75 800 EP 75 800 PRXR 75 800 R 75 802 75 802 AR 75 80...

Страница 4: ...GOW MAC Instrument Co 4 ...

Страница 5: ...to the bulk of the grains This leaves a clean surface to react with the impurities As the gas stream flows through the purifier contaminants are chemically sorbed at the surface of the purifier material By maintaining an operating temperature in the range of 350 450 C the impurities that have formed compounds on the surface can continuously diffuse into the bulk of the material thereby maintaining...

Страница 6: ... Watts 115V CAUTION IN ORDER TO MAINTAIN PROPER PURIFICATION EFFICIENCY DO NOT EXCEED ESTIMATED LIFETIME BEFORE REPLACING PURIFIER CARTRIDGE IF HELP IS NEEDED IN DETERMINING END OF LIFE CONTACT GOW MAC AT 610 954 9000 Each impurity removed to 10 ppb based on 10 ppm total inlet impurities at suggested maximum flow rate at 1 Atm operating pressure Higher flow rates can be used at higher operating te...

Страница 7: ...tainless steel tubing to a tank of Zero Grade Helium for Model 75 800 or Zero Grade Argon for Model 75 800 AR Attach the other end of tubing to INLET on the back of the Purifier using the provided VCR fitting B Attach the provided piece of clean 1 8 stainless steel with VCR fittings on both ends from the back of the Purifier OUTLET to CARRIER IN connection on the back of the GC or ZERO GAS IN for ...

Страница 8: ...grammed to do various tasks such as locking the keyboard restoring user settings etc ZONE Indicatesthecontroller zone 1 9 Advance Key Advances through parameter prompts Up and Down Keys In the Home page adjusts the set point in the lower display In other pages changes the upper display to a higher or lower value or changes a parameter selection Output 1 Indicator Light When lit it indicates Output...

Страница 9: ...he lower display Use the Up and Down keys to scroll through possible responses such as Clear CLr or Silence S L Then push the Advance or Infinity key to execute the action Er 1 Error Input 1 The sensor input generated a value lower than the allowable signal range Enter a valid input 100 Device Error Controller displays internal malfunction message at Power Up 7 Set Point Settings A To set the set ...

Страница 10: ... In general higher temperatures reduce outlet impurity levels for all gases except hydrogen Higher temperatures are recommended in case of high inlet impurity levels and or high flow rates iv The gas to be purified can now be allowed to flow through the unit at the desired flow rate CAUTION IF THE GAS LINE IS ACCIDENTLY VENTED TO AIR THE POWER SHOULD IMMEDIATELY BE TURNED OFF OTHERWISE THE PURIFIE...

Страница 11: ...res over 350 C and with a continued access to air The operator should place a VCR cap over the inlet outlet fittings for sufficient precaution before disposal It is then acceptable to deposit in an approved landfill in accordance with local state and or federal regulations When disposed of in this manner the waste poses no known environmental problems 10 Replacement Parts Part No Description 121 1...

Страница 12: ...W MAC Instrument Co 12 Contact GOW MAC for replacement parts GOW MAC Instrument Co 277 Brodhead Road Bethlehem PA 18017 U S A Tel 610 954 9000 Fax 610 954 0599 E mail sales gow mac com URL www gow mac com ...

Страница 13: ...y not covered by this warranty 6 Instrument parts which have been repaired or replaced during the warranty period are themselves warranted only for the remaining unexpired portion of the original one year warranty 7 Repairs adjustments and service performed after expiration of the one year warranty period shall be charged to the purchaser at the then current prices for parts labor and transportati...

Страница 14: ...GOW MAC Instrument Co 14 ...

Страница 15: ...n this instrument part Yes see below No proceed to D Please check the appropriate box Chemicals or Substances That Are Hazardous to Health Blood Body Fluids e g Urine Secretions Pathological Specimens Regulated Medical Wastes Infectious Substances or other Bio Agents e g Protein Enzymes Antibodies Radioactive Isotopes used in the area Detail type ECD Isotopic Labels etc and Activity in Micro Curie...

Страница 16: ...GOW MAC Instrument Co 16 ...

Отзывы: