GOW-MAC Instrument Co.
H. Zero Instrument
The detector signal reads the actual output signal to data collection (mV). Press “Zero” button
to zero the baseline signal. The instrument may require multiple attempts to bring the baseline
to ~0.00 mV.
I. Valve Control
1. Set the desired valve position for operation by pressing the valve buttons in “Valve
Control”
.
BLUE button indicates counter-clockwise (CCW) position and GREEN
button indicates (CW) position. See the supplied flow diagram for operating valve
configurations. The labels displayed on each valve may differ depending on the supplied
flow diagram.
2. “Remote” box allows operator to control the selected valves
REMOTELY
using the
software with contact closure controls controls. When the box is checked, each
respective
valve
(BLUE or GREEN) button is GRAYED out. See Section 7-4 for remote
valve
operation.
3. “Remote Lockout” locks the manual valve control function on the main screen to
remotely control the valves.
ALL
the valve buttons are GRAYED out when this option
is selected. To activate the “Remote Lockout” press “Menu” and enter code number “6”.
Press “Enter” to accept the setting, or “Esc” to exit to the main screen.
J. Save Settings
1. Customized settings and operating parameters can be stored for future operations.
Press “Menu” and enter code number “9” to save settings. Press “Enter” to accept or
“Esc” to exit to main screen.
K. Load Settings
1. To load the stored settings for the operation, Press “Menu” and enter code number “8”.
The operator can select “Default” to return all settings to factory or “Saved” to load the
customized settings. Select “Cancel” to return to the main screen. The set values on
the main screen will change based on the selected settings.
2.
The
DEFAULT
factory settings are as follows:
Column Temperature
: 70 °C
DID Temperture
: 40 °C
Start Voltage
: 750 V
Run Voltage
: 525 V
Polarization Voltage
: 160 V
Range
: 1
Gain
: 1x Low
Содержание 5900 Series