Gossen MetraWatt PROFITEST PVSUN Скачать руководство пользователя страница 7

GMC-I Messtechnik GmbH

7

10

Data Storage with PROFITEST PV

SUN MEMO

The 

PROFITEST PV

SUN MEMO

 test instrument is equipped with an 

internal memory for the storage of 10,240 data records. The 
stored data can be read out via a USB port for subsequent evalu-
ation from a PC. Furthermore, the included PC software allows for 
the creation of individual test structures.

10.1

Operating and Display Elements

1 LCD panel with background illumination

2 USB  interface

3 Sensor socket for temperature measurement (optional)

4 ON/OFF switch / background illumination

5 Low-resistance measurement 

R

LOW

 / measurement cable 

compensation 

COMP

6 Short-circuit measurement 

I

SC

 / insulation resistance 

measurement 

R

ISO

 per EN 62446

7 General insulation resistance measurement 

R

ISO

 / 

selection of insulation test voltage

8 Battery level indicator 

9 Selection of objects or strings

10 Switch between display / activation of storage function

11 Switch between objects and strings

12 Selection of objects or strings

13 Back: battery compartment

14 Measuring input ground terminal (

) / 

insulation measurement / continuity test

15 Measuring input 

 / insulation measurement / continuity test

16 Measuring input 

+

10.2

Function Description

Switch between standard measurement display 
and storage display

Switch to previous object or string

Switch to next object or string. If the last object or 
string is reached, a new object or a new string 
can be created by pushing this key (> 2 s).

Switch between objects and strings. Symbol 

 

indicates whether objects or strings are active 
(e. g.: 

OBJ:Object1).

Standard measurement 
display without memory 
function

The storage function is 
not activated before 
switch-over to storage 
display, see screenshot 
below.

Standard measurement 
display with activated 
storage function.

 

If a measurement is per-
formed, the measured 
values are stored under 
the object and string 
selected before.

Note

Stored values are overwritten in the activated object/
string.

Storage display

OBJ:Object1 / 

STR:String1 are active. 
Presently, one object (1/1) 
and one string (1/1) have 
been created.
Date / time

Measured values currently stored (meaning of abbreviations):
U

OC

Open-circuit voltage

I

SC

 Short-circuit 

voltage

R

ISO

 Insulation 

resistance

R

low

 Low-resistance 

measurement

E

 

Solar irradiation (optional with 

PROFITEST PV

SUN

-

SOR

)

T

 Module 

temperature (optional with 

PROFITEST PV

SUN

-

SOR

)

Inclination angle of module (optional with 

PROFITEST PV

SUN

-

SOR

)

I

b

 

Operating current (optional with current clamp)

(Alphanumeric) names 
can be assigned to 
objects and strings with 
the PC software
(e. g. 

OBJ:

 t

est client

).

Note

If objects and strings are extended in the test instrument, 
it is not possible to assign alphanumeric characters. 
There is only an incrementation instead (gradual increase 
by one character each).

2

1

16

15

14

4

5

6

7

13

8

3

9 10

11 12

Battery full

Battery weak

 U

OC

=0V 

 I

SC

=--- 

 R

ISO

=--- 

 U

OC

=0V 

 I

SC

=--- 

 R

ISO

=--- 

? OBJ:Objekt1                 (1/1) 
    STR:String1                  (1/1) 
    20.02.2012            10:00:00 
U

OC

=0V                 I

SC

=0.0A 

R

ISO

=>20M

         R

low

>10

 

E=---                     T=--- 

=---                      Ib=--- 

 

? OBJ:Testkunde             (1/1) 
    STR:String West           (1/1) 
    20.02.2012            10:00:00 
U

OC

=0V                 I

SC

=0.0A 

R

ISO

=>20M

         R

low

>10

 

E=---                     T=--- 

=---                      Ib=--- 

www.

.com

[email protected]

1.800.561.8187

Содержание PROFITEST PVSUN

Страница 1: ...Operating Instructions PROFITEST PVSUN and PVSUN MEMO Test Instruments for Testing PV Modules and Strings per DIN EN 62446 VDE 0126 23 3 349 672 15 11 3 16 www com information itm com 1 800 561 8187 ...

Страница 2: ...g instructions 1 USB interface cable PROFITEST PVSUN MEMO only 2 Significance of Symbols Meanings of Symbols on the Instrument Warning concerning a point of danger attention observe documentation Protection class II device continuous doubled or reinforced insulation This device may not be disposed of with the trash Fur EC mark of conformity Meanings of Symbols in the Operating Instructions Caution...

Страница 3: ...sured the instrument must be removed from service and secured against unintentional use Safety of the user is no longer assured if the instrument Demonstrates apparent damage No longer performs the desired measurements Has been stored for too long under unfavorable conditions Has been subjected to mechanical stressing during transport The instrument may not be opened dismantled or modified in any ...

Страница 4: ...terminal must also be tested Warning Disconnect all solar modules from the system before performing the measurement Compensation for Measurement Cable Resistance Ohmic measurement cable resistance can be subtracted from the measurement results automatically Proceed as follows in order to determine the resistance offset value Switch the PROFITEST PVSUN test instrument on Connect the measurement cab...

Страница 5: ...r module under test from the system before performing measurement Attention Insulation measurement per EN 62446 is executed be tween short circuited and terminals Connect the PROFITEST PVSUN test instrument to the module with the help of suitable solar plugs Connect the ground terminal at the PROFITEST PVSUN to the module frame Switch the PROFITEST PVSUN test instrument on The module s open circui...

Страница 6: ...ressing the RISO 250V 500V 1000V key The last selected value is saved after a short period of time and is then cleared from the display Example Executing the Insulation Resistance Measurement Press the RISO 250V 500V 1000V key and read the measured value from the display Example Note If the minimum permissible limit value is fallen short of 1 M or 0 5 M appears for RISO 9 6 Ground Fault Test If th...

Страница 7: ...ch between objects and strings Symbol indicates whether objects or strings are active e g OBJ Object1 Standard measurement display without memory function The storage function is not activated before switch over to storage display see screenshot below Standard measurement display with activated storage function If a measurement is per formed the measured values are stored under the object and stri...

Страница 8: ...lect directory C Programme PVSUN MEMO Control Center USB Driver unless you have changed the installation path Ignore the message regarding unknown provenance of the software Complete the driver installation Please check PVSUN MEMO Memory COMxx should be displayed in the device manager under Connections Start the installed program PVSUN MEMO Control Center Click on icon Search device 11 4 Creation ...

Страница 9: ...e 0 to 10 Test current 200 mA Resolution 0 1 Accuracy 1 rdg 1 d No of measurements Approx 500 low resistance measurements batteries 1 5 V per IEC LR6 Display LCD Backlit multiple display Dot matrix 128 x 64 pixels Reference Conditions Ambient temperature 23 C 2 K Relative humidity 40 to 75 Battery voltage 6 V 1 V Ambient Conditions Operating temperature 0 to 40 C Storage temperature 10 C to 60 C R...

Отзывы: