Introduction

These operating instructions contain information and directions 
which are necessary for safe operation and use of the instrument. 
Before using the instrument, the operating instructions must be 
carefully read and all points contained therein must be followed 
correctly. Non-observance of the operating instructions, as well 
as warnings and directions included therein, may result in severe 
injury to the user or damage to the instrument.

The 

PROFITEST PV

SUN

 is used for simple testing of solar cells and 

modules. The terms used in these operating instructions to desig-
nate systems for generating electrical power from sunlight include: 
solar system, photovoltaic system, PV system and the like.

Scope of Functions

Voltage measurement to 1000 V DC

Short circuit current measurement to 20 A DC

Selection of insulation test voltage: 250, 500 or 1000 V DC

Insulation measurement to 20 M

 with display of limit value

Testing of protective conductor continuity

Earth fault measurement

Polarity check

Backlit LCD panel

Optional temperature measurement

4

Transport and Storage

Please retain the original packaging for subsequent shipment of 
the instrument. Transport damages due to inadequate packaging 
are excluded from the guarantee.
The instrument may only be stored in dry, closed rooms.
If the instrument is subjected to extreme temperature during 
transport, it requires at least 2 hours for acclimatization before it is 
switched on.

5

Safety Precautions

The instrument has been manufactured and tested in accordance 
with IEC/EN 61010-1, “Safety requirements for electrical equip-
ment for measurement, control and laboratory use”, and has been 
shipped from the factory in flawless technical safety condition. In 
order to assure that this condition is retained, the user must 
observe the safety precautions included in these operating 
instructions.

Caution! Dangerous Voltage

In order to prevent electrical shock, corresponding safety 
precautions must be implemented when working with 
voltages of greater than 120 V (60 V) DC or 50 V (25 V) 
AC TRMS. In accordance with DIN VDE, these values 
specify the limit for exposed (touchable) voltages (the val-
ues in parentheses apply to, for example, the fields of 
medicine and agriculture).

Before each measurement, make sure that the measurement 
cables and the measuring instrument are in flawless condition. 
The measuring instrument may only be used within the specified 
measuring ranges.
If the safety of the user is no longer assured, the instrument must 
be removed from service and secured against unintentional use.

Safety of the user is no longer assured if the instrument:

Demonstrates apparent damage

No longer performs the desired measurements

Has been stored for too long under unfavorable conditions

Has been subjected to mechanical stressing during transport

The instrument may not be opened, dismantled or modified in any 
way. The instrument may only be used with recommended 
accessories. The use of unsuitable accessories is impermissible.
Trade association accident prevention regulations for electrical 
systems and operating equipment must be adhered to during all 
work with the instrument.
Do not allow the

 instrument to warm up

 due to exposure to direct 

sunlight. Flawless functioning and a long service life cannot other-
wise be assured.

Attention!

!

The 

internal warming of the test instrument

 is measured 

electronically. An inpermissible overheating is signalled 
by I

SC

 and 

°C 

in the display while further measurements 

are being disabled. Allow the test instrument to cool 
down before continuing measurements.

6

Use for Intended Purpose

The instrument may only be used under the conditions and for the 
purposes for which it has been designed. In this respect, special 
attention must be given to safety precautions, technical data 
regarding ambient conditions and use in a dry environment.

Warning!

The instrument may not be operated at measuring cate-
gory II, III and IV mains power circuits. 
In accordance with EN 61010-1:2010, the instrument 
has been designed without a rated measuring category, 
and may only be used at direct voltage circuits with up to 
1000 V without transient overvoltages.

Operating safety is no longer assured in the event of modification 
or alteration. The instrument may only be opened by authorized 
service technicians.
Never operate the instrument at a higher voltage than specified in 
the technical data! The instrument may otherwise be destroyed or 
permanently damaged.

7

Operating and Display Elements 

1

LCD panel with background illumination

2

Sensor socket for temperature measurement (optional)

3

On/Off switch / background illumination

4

Low-resistance measurement 

R

LOW

 / 

measurement cable compensation 

COMP

5

Short-circuit measurement 

I

SC

 /

insulation resistance measurement 

R

ISO

 per EN 62446

6

General insulation resistance measurement RISO /
selection of insulation test voltage

7

Measurement input 

+

 

8

Measurement input – / insulation measurement / continuity test

9

Measurement input ground terminal / insulation measurement / 
continuity test

10 Back: battery compartment
11 Battery level indicator 

2

1

7

8

9

3 4

5 6

10

11

Battery full

Battery weak

www.

.com

[email protected]

1.800.561.8187

Содержание PROFITEST PVSUN

Страница 1: ...Operating Instructions PROFITEST PVSUN and PVSUN MEMO Test Instruments for Testing PV Modules and Strings per DIN EN 62446 VDE 0126 23 3 349 672 15 11 3 16 www com information itm com 1 800 561 8187 ...

Страница 2: ...g instructions 1 USB interface cable PROFITEST PVSUN MEMO only 2 Significance of Symbols Meanings of Symbols on the Instrument Warning concerning a point of danger attention observe documentation Protection class II device continuous doubled or reinforced insulation This device may not be disposed of with the trash Fur EC mark of conformity Meanings of Symbols in the Operating Instructions Caution...

Страница 3: ...sured the instrument must be removed from service and secured against unintentional use Safety of the user is no longer assured if the instrument Demonstrates apparent damage No longer performs the desired measurements Has been stored for too long under unfavorable conditions Has been subjected to mechanical stressing during transport The instrument may not be opened dismantled or modified in any ...

Страница 4: ...terminal must also be tested Warning Disconnect all solar modules from the system before performing the measurement Compensation for Measurement Cable Resistance Ohmic measurement cable resistance can be subtracted from the measurement results automatically Proceed as follows in order to determine the resistance offset value Switch the PROFITEST PVSUN test instrument on Connect the measurement cab...

Страница 5: ...r module under test from the system before performing measurement Attention Insulation measurement per EN 62446 is executed be tween short circuited and terminals Connect the PROFITEST PVSUN test instrument to the module with the help of suitable solar plugs Connect the ground terminal at the PROFITEST PVSUN to the module frame Switch the PROFITEST PVSUN test instrument on The module s open circui...

Страница 6: ...ressing the RISO 250V 500V 1000V key The last selected value is saved after a short period of time and is then cleared from the display Example Executing the Insulation Resistance Measurement Press the RISO 250V 500V 1000V key and read the measured value from the display Example Note If the minimum permissible limit value is fallen short of 1 M or 0 5 M appears for RISO 9 6 Ground Fault Test If th...

Страница 7: ...ch between objects and strings Symbol indicates whether objects or strings are active e g OBJ Object1 Standard measurement display without memory function The storage function is not activated before switch over to storage display see screenshot below Standard measurement display with activated storage function If a measurement is per formed the measured values are stored under the object and stri...

Страница 8: ...lect directory C Programme PVSUN MEMO Control Center USB Driver unless you have changed the installation path Ignore the message regarding unknown provenance of the software Complete the driver installation Please check PVSUN MEMO Memory COMxx should be displayed in the device manager under Connections Start the installed program PVSUN MEMO Control Center Click on icon Search device 11 4 Creation ...

Страница 9: ...e 0 to 10 Test current 200 mA Resolution 0 1 Accuracy 1 rdg 1 d No of measurements Approx 500 low resistance measurements batteries 1 5 V per IEC LR6 Display LCD Backlit multiple display Dot matrix 128 x 64 pixels Reference Conditions Ambient temperature 23 C 2 K Relative humidity 40 to 75 Battery voltage 6 V 1 V Ambient Conditions Operating temperature 0 to 40 C Storage temperature 10 C to 60 C R...

Отзывы: