24
GMC-I Messtechnik GmbH
Electrical Safety
Protection class
II
Nominal line voltage
230 V
Test Voltage
Mains + PE (mains) + 2 mA socket for testing for absence of voltage
at test socket, connector sockets for phase and protective
conductors, as well as clip: 3 kV
∼
mains to PE (mains) + 2 mA socket
1.5 kV
∼
Measuring category
II
Pollution degree
2
Safety shutdown
If test instrument overheats
Electromagnetic Compatibility (EMC)
Product standard
EN 61326-1: 2006
Ambient Conditions
Operation
−
10 ... + 55
°
C
Storage
−
25 ... + 70
°
C
Humidity
Max. 75%, no condensation allowed
Elevation
To 2000 m
Mechanical Design
Dimensions
W x H x D: 190 x 140 x 95 mm
Weight
1.3 kg
Protection
Housing: IP 40, connections: IP 20
Table Excerpt Regarding Significance of IP Codes
Interference Emission
Class
EN 55022
A
Interference Immunity
Test Value
Feature
EN 61000-4-2
Contact/atmos. – 4 kV / 8 kV B
EN 61000-4-3
10 V/m
B
EN 61000-4-4
Mains connection – 2 kV B
EN 61000-4-5
Mains connection – 1 kV A
EN 61000-4-6
Mains connection – 3 V B
EN 61000-4-11
0.5 period / 100%
A
IP XY
(1
st
char. X)
Protection against
foreign object entry
IP XY
(2
nd
char. Y)
Protection against the
penetration of water
0
not protected
0
not protected
1
≥
50.0 mm diameter
1
vertically falling drops
2
≥
12.5 mm diameter
2
vertically falling drops with
enclosure tilted 15°
3
≥
2.5 mm diameter
3
spraying water
4
≥
1.0 mm diameter
4
splashing water