12
Gossen Metrawatt GmbH
8
Technical Data
Article Number
Z745S
Residual Current Measuring Function
Measuring Range
0
20 mA
Transformation Ratio
1 V per 10 mA
Intrinsic Uncertainty
(5% of rdg. + 0.05 mA)
Nominal Ranges of Use
Line Voltage
L1/L2/L3/N
207
253 V AC
Frequency
49
51 Hz
Temperature
0 °C
+40 °C
Line Voltage
Waveshape
sinusoidal
Reference Conditions
Ambient Temperature
+23
C
2 K
Relative Humidity
50%
5%
Line Voltage
230 V/400 V
10%
Frequency
of Measured Quantity
50 Hz
0.2%
Ambient Conditions
Operating Temperatures –10
+ 40
C
Storage Temperatures
–25
+ 60
C
Humidity
max. 75%, no conden
-
sation allowed
Elevation
up to 2000 m
Power Supply
Nominal Line Voltage
3~230/400 V/50 Hz/CAT II
Connection only permissible with overload protec
-
tion device I
n
= 16 A I
2
1.45 I
n
Electrical Safety
Pollution degree
2
Safety Class
I per DIN EN 61140/
VDE 0140-1
Residual Current
Shutdown (4-pole) at
I
dN
= 14,6 mA,
t
a
= 82 ms
Device Fuses
F0315 L250V
5 x 20
T32mA L 250 V
DIN EN 60127-2
Electromagnetic Compatibility
Intrinsic Uncertainty
EN 61326-1:2013 class B
Intrinsic Uncertainty
EN 61326:2013
Mechanical Design
Protection
case: IP40
terminals: IP20
Extract from table on the meaning of IP codes
Dimensions
405 x 300 x 220 (mm)
with lid
Weight
approx. 6.7 kg
Note!
Observe the technical data of the respective
test instrument.
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2
nd
digit Y)
Protection against the
penetration of water
0
not protected
0
not protected
1
50.0 mm dia.
1
vertically falling
drops
2
12.5 mm dia.
2
vertically falling
drops with enclo
-
sure tilted 15
3
2.5 mm dia.
3
spraying water
4
1.0 mm dia.
4
splashing water