Good Will Instrument PSS-2005 Скачать руководство пользователя страница 5

    PSS-3203/2005 PROGRAMMABLE POWER SUPPLY

USER MANUAL

  4  

Interface

RS232 (GPIB interface option)

Power Source

AC100V, 120V, 220V

±

10%, 230V +10%/-6% 50/60Hz

Power
Consumption

PSS-2005: 215W
PSS-3203: 190W
Dimensions

110(W)

×

140(H)

×

320(D) mm.

Mechanical
Spec.

Weights

4.8 kg

Operation
Environmental

Indoor use
Altitude up to 2000 m
Ambient temperature :
    To satisfy specifications : 10

 to 35

 ( 50

°

 F to 95

°

F )

    Maximum operating ranges: 0

 to 40

( 32

°

F to 104

°

F )

Relative humidity: 85% RH(max.) non condensing
Installation Category: II
Pollution degree: 2

Storage
Temperature &
Humidity

-10

°

 to 70

, 70%RH (maximum)

Accessories

Power cord….............……….. 

×

  1

Instruction manual……………..

×

  1

Programmer manual................. 

×

  1

Test Lead……….………………

×

  1

    PSS-3203/2005 PROGRAMMABLE POWER SUPPLY

USER MANUAL

  5  

3.PRECAUTIONS BEFORE OPERATION

3-1.Unpacking the Instrument

The product has been fully inspected and tested before shipping from the

factory. Upon receiving the instrument, please unpack and inspect it to

check if there is any damage caused during transportation. If any sign of

damage is found, notify the bearer and/or the dealer immediately.

3-2.Checking the Line Voltage

The product can be applied by any kind of line voltages shown in the table

below. Before connecting the power plug to an AC line outlet, make sure

the voltage selector of the rear panel is set to the correct position

corresponding to the line voltage. It might be damaged the instrument by

connecting to the wrong AC line voltage.

WARNING.  To avoid electrical shock the power cord

protective grounding conductor must be connected to

ground.

AVERISS: Pour éviter les chocs électriques, le fil de terre

du cordon secteur doit impérativement être relié à la

terre.

When line voltages are changed, replace the required fuses shown as below:

Model

Line

voltage

Range

Fuse

Line

voltage

Range

Fuse

PSS-3203

PSS-2005

100V
120V

90-110V

108-132V

T3A

250V

220V
230V

198-242V
216-250V

T1.6A

250V

Содержание PSS-2005

Страница 1: ...12 13 13 13 6 MAINTENANCE 6 1 Fuse Replacement 6 2 Line Voltage Conversion 6 3 Adjustment and Calibration 6 4 Cleaning 14 14 14 15 18 7 THE SYSTEM DIAGRAM AND DESCRIPTION 7 1 Block Diagram 7 2 The Ope...

Страница 2: ...loured Blue or Black The wire which is coloured Brown must be connected to the terminal marked with the letter L or P or coloured Brown or Red If in doubt consult the instructions provided with the eq...

Страница 3: ...0 4 4 1995 Surge Immunity EN 61000 4 5 1995 Conducted Susceptibility EN 61000 4 6 1996 Voltage Dips Interrupts EN 61000 4 11 1994 Safety Low Voltage Directive 73 23 EEC amended by 93 68 EEC IEC EN 610...

Страница 4: ...0 21V 1 Voltage 3mV 5mV rating current 3 0A Load Effect Current 3mA 5mA rating current 3 0A Voltage 3mV Source Effect Current 3mA Voltage 10mV Current 1mA 2mA rating current 3 0A Resolution OVP 10mV V...

Страница 5: ...ng from the factory Upon receiving the instrument please unpack and inspect it to check if there is any damage caused during transportation If any sign of damage is found notify the bearer and or the...

Страница 6: ...32 to 104 F To operate the instrument exceeding this specific temperature range may cause damage to the circuits of instrument Do not use the instrument in a place where strong magnetic or electric f...

Страница 7: ...for value input or setting confirmation 8 F C The knob for switching over coarse and fine adjustment 9 MENU The category of function setting Output OVP OCP Contrast Buzzer Interface PS After switching...

Страница 8: ...RAMMABLE POWER SUPPLY USER MANUAL 11 Example If want to set the current as 1 234A first using the F C knob to switch the cursor to 1mA step and adjust the value to 34 then switch the cursor to 100mA s...

Страница 9: ...indow will be appeared then using the wheel knob to modify the value and press ENTER to complete the setting Note The system will detect the interface used at present automatically and switch the dete...

Страница 10: ...tage to another is done by change AC selects switch The rear panel identifies the line voltage to which the unit was factory set To convert to a different line voltage perform the following procedure...

Страница 11: ...e Max measured by the DMM with the knob and press ENTER During the current value input use the PSS 3203 2005 PROGRAMMABLE POWER SUPPLY USER MANUAL 17 F C button to switch over the input of 100mA range...

Страница 12: ...may leak into the cabinet and cause damage Do not use chemicals containing benzine benzene toluene xylene acetone or similar solvents Do not use abrasive cleaners on any portion of the instrument PSS...

Страница 13: ...materials can use the Vr to make adequate adjustment PSS 3203 2005 PROGRAMMABLE POWER SUPPLY USER MANUAL 21 The operation in C C Mode is similar to that in C V Mode The MPU will send 2100 counts 2 1A...

Отзывы: