Golander
BT103S Manual
14
counterclockwise direction.
Note
: The external DC power source can be 5V or 12V. If it is
24V, 1.5K resistor is needed to protect internal circuit.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
ADC_W
AGND
0-5V/0-10V
/4-20mA
24VDC
VCC_W
COM
RS_W
CW_W
1.5K
1.5K
Figure 15. DB15 Wiring with External 24VDC Power Source
7.6
Time Dispense Mode
Pump will dispense fluid automatically by setting duration for each
dose, pause time between doses and number of cycles. When
dispensing, the display will show the dispensing time or lag time
that has passed (1.2s on the figure below), and the total cycles (4
on the figure) that has dispensed.
Disp
32.5rpm
Disp
32.5rpm 1.2s 4
Dispense Stopped
Dispense Running
Figure 16. Time Dispense Mode
To set Time Dispense mode:
1) Turn on the power switch. Pump will display the main screen.
2) Press
MODE
key to change the mode to Time Dispense
mode (Disp shown on the screen).
3) Press
MENU
key to enter the Setting menu.