www.g-mw.de
7
Repair work and modifications to the test device must only be carried out by the
manufacturer itself or specialists authorised by the manufacturer!
Only specialists may perform repair work on mains adapter cables.
Only the original replacement parts specified by the manufacturer may be fitted
and used!
If safe operation is no longer possible, e.g. due to:
-
Visible damage,
-
Improper storage,
-
Improper transportation,
-
Failure of a phase control lamp,
-
Failure of measurement functions, etc.,
the test device must no longer be operated! The test device must be
decommissioned without delay and secured to prevent unintentional commissio-
ning! The test device may be repaired by centres authorised by the manufacturer!
Measurements may only be performed on protective earth conductor resistances
and touch currents under certain circumstances
and in compliance with the corresponding hazard warnings!
Note that the test objects may carry high voltages, e.g. from charged capacitive
circuits!
Do not connect the test object to the test socket until the mains connection is
technically safe!
WARNING!
When connecting the test object to a test socket, voltages dangerous
to the touch can result on a defective test object or touchable conductive parts
not connected to the protective earth conductor.