GASURVEYOR 500 USER HANDBOOK
B-6
4) Turn the regulator valve counter clockwise to open
the valve slightly. Make sure that the gas is flowing
before connecting the sample tubing to the
instrument, otherwise an instrument sample fault
may occur.
5) Connect tubing from regulator to instrument inlet
then adjust the regulator valve to maintain a
constant flow of gas (counter clockwise to
increase flow and clockwise to decrease).. The
correct flow rate is achieved when the ball in the
indicator tube floats just above its resting position.
6) Wait for the instrument gas reading to settle.
7) If the displayed reading corresponds to the
concentration of calibration gas, i.e. 50% LEL
(2.5% Methane in Air), proceed to paragraph 10.
8) If the displayed reading does not correspond to
the concentration of calibration gas, i.e. 50% LEL
(2.5% Methane in Air), press and hold Switch One
(
) to enter SPAN mode.
SPAN mode is indicated by the selected range, in
this case LEL, and SPN alternating in the display
as shown in Figure B-6.
Figure B-6 SPAN Mode Display
Содержание Gasurveyor 500 Series
Страница 1: ...Gasurveyor 500 User Handbook Distributed By 713 844 1300 info heathus com www heathus com...
Страница 6: ...GASURVEYOR 500 USER HANDBOOK iv...
Страница 8: ...GASURVEYOR 500 USER HANDBOOK...
Страница 14: ...GASURVEYOR 500 USER HANDBOOK xii...
Страница 18: ...GASURVEYOR 500 USER HANDBOOK 1 4...
Страница 31: ...3 3 OPERATION Figure 3 2 CSM Warm Up Note Battery status is displayed as either OK or LO at start up...
Страница 54: ...GASURVEYOR 500 USER HANDBOOK 6 4...
Страница 56: ...GASURVEYOR 500 USER HANDBOOK 7 2...
Страница 60: ...GASURVEYOR 500 USER HANDBOOK A 4...
Страница 70: ...GASURVEYOR 500 USER HANDBOOK B 10...
Страница 96: ...C 26 GAS MEASUREMENT INSTRUMENTS LTD...
Страница 101: ......