GASURVEYOR 500 USER HANDBOOK
B-4
In CAL mode, the instrument switches have the functions
shown in Figure B-4.
–
–
Figure B-4 FCM Switch Functions
Zeroing the Instrument
1) Enter FCM. See the previous section ENTERING
FCM.
2) Double press Switch Two (
) to zero the
current gas range.
3) Press Switch Two (
) to select the next gas
range.
Note: Each press of Switch Two (
) selects the next
gas range to be zeroed until finally the Cal Due
Date is displayed. Editing the Cal Due Date feature
is explained later in this procedure.