Gasurveyor 500

User Handbook

Distributed  By:

713-844-1300

[email protected]

www.heathus.com

Содержание Gasurveyor 500 Series

Страница 1: ...Gasurveyor 500 User Handbook Distributed By 713 844 1300 info heathus com www heathus com...

Страница 2: ...e in helping us to improve our customer publications Please send any comments that you have to our Sales Department at GMI Contact details are provided inside the back cover of this handbook Instrumen...

Страница 3: ...operational differences between the latest product and this manual This Handbook is an important part of the Gasurveyor 500 product Please note the following points It should be kept with the instrum...

Страница 4: ...ents are known or suspected it is recommended that more frequent response checks are carried out The chemical compounds that can cause loss of sensitivity include Silicones Lead Halogens and Sulphur D...

Страница 5: ...iii DECLARATION OF CONFORMITY Refer to current Declaration of Conformity document Part No 42907 supplied with product...

Страница 6: ...GASURVEYOR 500 USER HANDBOOK iv...

Страница 7: ...iger clarification and effect of CN 2313 17 07 03 4 Revised to include effect of CN 2291 26 07 04 5 Revised to include Appendix C translations 08 06 07 6 Revised to include Marine Equipment Directive...

Страница 8: ...GASURVEYOR 500 USER HANDBOOK...

Страница 9: ...LADVICE i SAFETY ii AREAS OF USE ii STORAGE HANDLINGANDTRANSIT ii DECLARATION OF CONFORMITY iii REVISION RECORD v INTRODUCTION 1 1 GENERALINFORMATION 2 1 Ranges of Operation 2 1 LEL 0 to 100 2 1 Volum...

Страница 10: ...osable Alkaline LR20 Dry Cell Batteries 2 9 Rechargeable NiCd Battery Pack 2 9 Filters 2 10 Liquid Crystal Display LCD 2 10 Before Use Checks 2 10 OPERATINGINSTRUCTIONS 3 1 Operating Modes 3 1 Confine...

Страница 11: ...ff 3 9 Changing Range 3 10 Activate the Audible Geiger Indication PPM range 3 10 Zeroing the PPM Range 3 10 Manual Datalogging Default 3 11 Summary of CGI Button Operation 3 11 Operator Messages Fault...

Страница 12: ...ALINFORMATION 7 1 Training 7 1 World Wide Web 7 1 TYPICALOPERATING PARAMETERS A 1 Size A 2 Weight A 2 Operating Temperature A 2 Humidity A 2 Construction A 2 Display A 2 Sampling System A 2 Response T...

Страница 13: ...on Procedure B 5 Edit Cal Due Date B 8 Quit And Save Changes B 9 Quit Without Saving Changes B 9 QUICK OPERATING INSTRUCTIONS C 1 English C 2 L italiano Italian C 6 Espa ol Spanish C 10 Svensk Swedish...

Страница 14: ...GASURVEYOR 500 USER HANDBOOK xii...

Страница 15: ...histicated four gas multi tasking instrument The instrument is designed to the latest standards and is certified for use in Hazardous Areas Figure 1 1 Gasurveyor 500 series Instrument The Gasurveyor 5...

Страница 16: ...e have provided in the following pages is the standard default of how the instrument would generally be configured with the possible options detailed in italic text GMI recommend that you take the tim...

Страница 17: ...y view the display The signals which consist of an audible beep and a visual pulse of the LED approximately every eight seconds are active in CSM mode and also active in CGI mode but for the PPM Flamm...

Страница 18: ...GASURVEYOR 500 USER HANDBOOK 1 4...

Страница 19: ...range is displayed in the top right corner of the display as LEL From 0 to 10 LEL the digital display resolves to 0 1 LEL From 10 to 100 LEL the digital display resolves to 1 LEL The analogue bar gra...

Страница 20: ...The digital display resolves the signal to 1 GAS with the analogue bar graph following in steps of 4 Figure 2 2 shows the Volume Gas display The detection principle for the Volume Gas range is therma...

Страница 21: ...only to significant levels The ppm range has no threshold set by default If your instrument has a threshold write the level here The ppm range has the ConfidenceAudible and Confidence Visual options a...

Страница 22: ...levels below the threshold as shown in Figure 2 4 Figure 2 4 Threshold Display Option Oxygen 0 25 This range displays the percentage Oxygen content of the sample The instrument range is displayed in t...

Страница 23: ...Section 3 The audible alarm is rated up to 85 dB A The visual alarm is a red LED indicator which protrudes from the instrument top plate allowing viewing from any angle above the top plate The instrum...

Страница 24: ...ction default is that the alarms will be latching and both the audible and visual functions will operate That is where an alarm level is exceeded the instrument LED flashes and the sounder pulses rapi...

Страница 25: ...priority alarm is displayed From highest to lowest the alarm priorities are Oxygen Volume Gas LEL Toxic Ranges When a toxic alarm is raised the priorities are Instantaneous Short Term TWA STEL Long Te...

Страница 26: ...utomatic and manual datalogging to be disabled when operating in CSM mode CGI Mode When the instrument is operating in CGI mode the datalogging is only by manual operation of Switch Three There is no...

Страница 27: ...to be used until it switches off automatically Rechargeable NiCd Battery Pack The GMI rechargeable battery pack provides approximately nine hours operational life from fully charged at ambient tempera...

Страница 28: ...iquid Crystal Display LCD The LCD shows the current gas readings in both analogue and digital form together with operational and status information The display is protected by a toughened glass cover...

Страница 29: ...on For Confined Space Monitor operation the instrument must be switched on by pressing and holding Switch One for approximately one second During warm up in CSM mode the instrument displays the prese...

Страница 30: ...mosphere to which the operator is exposed and generates alarms for low oxygen high oxygen high flammable gas high toxic gas Short Term TWA STEL and Long Term TWA LTEL Switching On in CSM Mode Pressing...

Страница 31: ...3 3 OPERATION Figure 3 2 CSM Warm Up Note Battery status is displayed as either OK or LO at start up...

Страница 32: ...eyor 500 instruments is twelve 12 months This period can be altered as an option however you should ensure that the instrument is always within its calibration period prior to use An option which ensu...

Страница 33: ...splay Options The Gasurveyor 500 instruments can be enabled to allow the user to view the minimum and maximum readings for each range during the instrument s operation since switch on The instrument a...

Страница 34: ...r an alarm if latching or acknowledge the alarm for 60 seconds if non latching Double pressing will only cancel the alarm when the gas level is below the threshold set in the instrument If more than o...

Страница 35: ...atic datalogging is enabled by default The PPM range is not available in CSM mode Summary of default CSM Button Operation f f O s e h c t i w S t n e m u r t s n I s e h c t i w S n O t n e m u r t s...

Страница 36: ...d in CGI mode as the instrument is generally used for measuring gas levels rather than monitoring for the presence of gas Switching On in CGI Mode Pressing and holding Switch Two turns the instrument...

Страница 37: ...r all Gasurveyor 500 instruments is twelve 12 months This period can be altered as an option however you should ensure that the instrument is always within its calibration period prior to use An optio...

Страница 38: ...ctivate the Audible Geiger Indication PPM range The audible Geiger indication can be activated by pressing and holding Switch Four A further press and hold of Switch Four mutes the indication Zeroing...

Страница 39: ...n in the top right hand corner of the display Pressing any but ton cancels this automatic switch off The instru ment will then allow another 30 minutes of use This 30 minute period is set as a default...

Страница 40: ...th the instrument s flow due to the sample path being blocked water ingress a blocked filter or pump failure In CGI mode the pump stops automatically The sample line filters etc should be checked for...

Страница 41: ...battery manufacturer temperature conditions usage etc With rechargeable batteries the BAT flag indicates approximately 30 minutes operation left As the battery power continues to fall the LCD flashes...

Страница 42: ...splayed if the measurement falls below 99 incorrect zero setting in either ppm or the toxic ranges 1 This message which can also appear after power on indicates that a calibration data error has been...

Страница 43: ...INTENANCE Rechargeable Battery Pack Three battery chargers are available from GMI a Standard Charger a Flatbed Charger and a Smart Charger Standard Charger The GMI Standard Charger takes approximately...

Страница 44: ...L mode which takes approximately 14 hours to charge a flat battery The Mode Select Switch can then be set to STAND BY where a trickle charge will maintain the battery in a fully charged state of readi...

Страница 45: ...he rechargeable pack should be fully discharged and charged on a regular basis of at least every three months The Smart Charger has the option of switching to discharge and fast charge cycle to provid...

Страница 46: ...an be done by running the instrument continuously or using the battery conditioning facility on the Smart Charger The batteries can be charged on the instrument but the instrument itself should be swi...

Страница 47: ...base 6 Replace battery cover and fasten base screws 7 Check that the instrument switches on and works to specification Filter Replacement Hydrophobic and cotton particulate filters in the probe handle...

Страница 48: ...r Probe Handle Assembly 2 Remove the cotton particulate filter and discard 3 Remove the hydrophobic filter 4 Clean the probe handle to make sure that it is free from dirt and water 5 Fit a new cotton...

Страница 49: ...ument can be linked to a PC running GMI Manual Calibration software Automatic Calibration The GMIAuto Test Calibration System allows calibration without manually changing gas cylinders The system link...

Страница 50: ...xpected This is no guarantee however as the precise application of the product is unknown to GMI Individual codes of practice may dictate shorter periods Regular checking establishes a pattern of reli...

Страница 51: ...Only 42744 Flatbed Charger c w Universal Power Supply 42124 Flatbed Charger excl Power Supply 12444 Universal Power Supply Standard Flatbed Charger 42110 Flatbed Charger Strap 13100 Smart Charger c w...

Страница 52: ...2ft 6ins approx 12287 Stainless Steel Probe Open End High Temp Applications 50cm 1ft 8ins approx 12688 Sample Line Adaptor 12895 Barbed Probe Solid End General Survey 69cm 2ft 3ins approx 12894 Barbe...

Страница 53: ...777 Filter Probe Assy Sampling Connector 1m Conductive Sample Line and Probe 13774 15m 48ft 9ins Conductive Sample Line c w Sampling Connector Filters 13792 35m 113ft 9ins Conductive Sample Line c w S...

Страница 54: ...GASURVEYOR 500 USER HANDBOOK 6 4...

Страница 55: ...raining Training courses are available on all our products Contact our Marketing Department for further details Tel 44 0 141 812 3211 Fax 44 0 141 812 7820 e mail sales gmiuk com World Wide Web Visit...

Страница 56: ...GASURVEYOR 500 USER HANDBOOK 7 2...

Страница 57: ...olume 0 to 100 1 2 1 1 Gas Gas Oxygen 0 to 21 0 1 0 5 0 5 3 reading 21 to 25 1 N A 0 5 3 reading Hydrogen 0 to 200 ppm 1 ppm H2 S 2 ppm 2 ppm 10 reading Sulphide Carbon 0 to 1000 ppm 1 ppm CO 15 ppm 1...

Страница 58: ...1000 or 0 10000 4 digit digital display 3 character range indication Operational flags Sampling System Integral pump with flow fail sensor Nominal flow rate is 0 5 to 0 7 litres per minute Typical flo...

Страница 59: ...ICAL OPERATING PARAMETERS Power Source 4 D size alkaline cells providing approximately 15 hours runtime at 20 o C 68 o F Rechargeble NiCd battery pack providing approximately 9 hours runtime at 20 o C...

Страница 60: ...GASURVEYOR 500 USER HANDBOOK A 4...

Страница 61: ...libration that should be noted The gas should be of known traceable quality and have total analysis The gas should be applied in the same manner as the instrument is used e g at a known pressure which...

Страница 62: ...o simplify switch operation when calibrating the instrument an overlay card shown in Figure B 2 is available and can be placed over the top face of the instrument to identify calibration switch functi...

Страница 63: ...nt switch off While OFF is displayed in the LCD and before the instrument actually switches off enter the four switch access code Note Allow at least one second between switch presses when entering th...

Страница 64: ...e Instrument 1 Enter FCM See the previous section ENTERING FCM 2 Double press Switch Two to zero the current gas range 3 Press Switch Two to select the next gas range Note Each press of Switch Two sel...

Страница 65: ...s 2 Make sure that the instrument pump is running and the gas range selected is compatible with the calibration gas Note A single press of Switch One toggles the pump Off On 3 Remove the cap from cali...

Страница 66: ...decrease The correct flow rate is achieved when the ball in the indicator tube floats just above its resting position 6 Wait for the instrument gas reading to settle 7 If the displayed reading corres...

Страница 67: ...ion The display may jump above and below required reading momentarily as the instrument performs the calibration Note If for any reason you require to exit SPAN mode without calibration of the instrum...

Страница 68: ...12 for each range to be calibrated otherwise edit Cal Due Date then Quit FCM as detailed in the following paragraphs Edit Cal Due Date 1 After successful zero calibration of all ranges press Switch T...

Страница 69: ...y reason you require to exit SPAN mode without changing Cal Due Date press and hold Switch Three 5 The instrument display will now show revised DUE date alternating with CAL Quit And Save Changes 1 Pr...

Страница 70: ...GASURVEYOR 500 USER HANDBOOK B 10...

Страница 71: ...peration guide for the Gasurveyor 500 series instrument Each language and pages reference is as follows English pages C 2 to C 5 L italiano Italian pages C 6 to C 9 Espa ol Spanish pages C 10 to C 13...

Страница 72: ...1ATEX2292 II 2 G UL Class 1 Groups A B C and D This equipment is designed and manufactured to protect against other haz ards as defined in paragraph 1 2 7 of Annex II of the ATEX Directive 94 9 EC Any...

Страница 73: ...n the instrument for servicing BAT Indication that the batteries will soon require replacement At this point there will be approximately 60 minutes left in a set of alkaline batteries although this fi...

Страница 74: ...s initiates the instrument s warm up cycle followed by instrument type and software version then battery status as shown If the instrument is switched on in CSM mode all default alarm levels are then...

Страница 75: ...le press of Switch Three performs a manual datalog Confirmation is via the STORE flag being displayed briefly and the buzzer sounding Automatic Datalogging CSM only A double press of Switch Three togg...

Страница 76: ...fiammabile nell atmosfera La serie Gasurveyor 500 certificato come EEx iad IIC T4 20o C Tamb 50o C BAS01ATEX2292 II 2 G UL Class 1 Groups A B C and D Questo equipaggiamento progettato e costruito per...

Страница 77: ...i calibrazione Lo strumento deve essere spento e riacceso in aria pulita Se il guato permane inviarlo al Servizio Assistenza BAT Indica che le batterie dovrebbero al pi presto essere sostituite A ques...

Страница 78: ...pulsante due per per accendere lo strumento e la pompa Questo dar inzio al ciclo di riscaldamento Tutti i segmanti del LCD si illuminano segue sul Display LCD il tipo di strumento e la versione del S...

Страница 79: ...uale Una singola pressione del tasto tre permette il datalogging manuale La conferma il messaggio STORE sul display e un bip acustico Datalogging Automatico solo per la modalit CSM Una doppia pression...

Страница 80: ...flamable y no se debe abrir en presencia de una atm sfera explosiva Los instrumentos Gasurveyor 500 cumplen con EEx iad llC T4 20o C Tamb 50o C BAS01ATEX2292 II 2 G UL Class 1 Groups A B C y D Este eq...

Страница 81: ...de gas Apague el equipo y reinicie en aire limpio ZERO FAULT Error cero Indicaci n donde el cero est fuera de los l mites de calibraci n Apague el instrumento y reinicie en aire limpio Si el error no...

Страница 82: ...ia el ciclo de warm up Todos los elementos se mostrar n en el display Seguido por el modelo del instrumento y la versi n del software Alternate flash Despu s el instrumento indica la fecha de calibrac...

Страница 83: ...uardar la medida manualmente Confirmaci n a trav s de la se al STORE y la sirena suena Datalogging autom tico s lo CSM Pulse el Bot n Tres dos veces para activar o desactivar el datalogging Si est act...

Страница 84: ...ags normala rutiner f ljas M tkammaren f r br nnbar gas r en Ex klassad enhet och f r ej ppnas d risk f r s dan gas f religger Gasurveyor serie 500 instrument r klassat enligt EEx iad IIC T4 20o C Tam...

Страница 85: ...gger utanf r sin gr ns St ng av instrumentet och terstarta det i ren luft Om felet ej f rsvinner l mnas instrumentet f r service till kvalificerad personal BAT Betyder att batteriet snart beh ver byta...

Страница 86: ...knapp 2 f r att s tta p instrumentet och pumpen Detta startar en automatisk kontroll och uppv rmning av instrumentet Alla tecken visas i displayen och tf ljs av Instrumentets beteckning mjukvaruversi...

Страница 87: ...det Detta bekr ftas av STORE indikering och en kort ljudsignal Automatisk dataloggning endast i CSM l ge Dubbeltryck p knapp 3 startar stoppar automatisk loggning av m tv rdena STORE indikering visar...

Страница 88: ...ure som din organisation har foreskrevet Forbr ndings kammer er brandsikker tilbeh r og m ikke bnes i almindelig atmosf re Ethvert krav i forbindelse med produkt ansvar eller f lge skade p tredje part...

Страница 89: ...nul gr nsen er uden for kalibrerings omr det Sluk instrumentet og t nd igen i frisk luft Hvis fejlen ikke er v k send instrumentet til service BAT Fort ller at batteriet snart l ber t r for str m Alt...

Страница 90: ...g hold nede den knap 1 s startes opvarmnings processen og pumpen Alternate flash Under opvarmning identificeres model serienummer software version batteristatus kalibrerings m ned og r Herefter vil di...

Страница 91: ...p knap 3 starter datalogging STORE vises p displayet og en summe lyd lyder Automatisk datalogging CSM 2 tryk p knap 3 starter stopper automatisk datalogging N r den er t ndt vises STORE p displayet Sl...

Страница 92: ...eer Gasurveyor 500 series instrumenten zijn gecertifieerd zoals EEx iad IIC T4 20o C Tamb 50o C BAS01ATEX2292 II 2 G UL Klasse 1 Groepen A B C en D Dit toestel is ontwikkeld en gemaakt om ons te besch...

Страница 93: ...unt buiten de kalibratielimieten valt Schakel het instrument uit en opnieuw aan in frisse lucht Indien de foutmelding niet verdwijnt breng het toestel terug voor onderhoud BAT Indicatie dat de batteri...

Страница 94: ...oor opwarming Alle LCD segmenten worden en daarna de batterij status zoals getoond Indien het instrument aangeschakeld wordt in CSM worden alle standaard alarmniveau s getoond Dan toont het instrument...

Страница 95: ...ordt door de STORE boodschap kort getoond en de buzzer maakt lawaai Automatische Datalogging enkel CSM Een dubbele druk op knop 3 zet de automatische datalogging Aan en UIT Indien AAN wordt de STORE b...

Страница 96: ...C 26 GAS MEASUREMENT INSTRUMENTS LTD...

Страница 97: ...on 3 7 3 11 C Cal Due Date B 8 CALIBRATION 5 1 B 1 Calibration Date Features 3 4 3 9 Calibration Procedure B 5 CALIBRATION VALIDITY 6 4 Calibration Validity 5 2 6 2 CGI 3 1 3 8 3 11 CGI Mode 2 8 CGI m...

Страница 98: ...EEE 3 13 EN 50104 A 2 English xi C 1 Entering FCM B 3 Espa ol Spanish C 1 F Fault Flags 3 12 FCM B 1 B 3 features 1 2 FIELD CALIBRATION B 1 C 1 Field Calibration Procedure B 5 Filter Replacement 4 5 F...

Страница 99: ...S A 1 Power Source A 3 ppm Flammable 2 2 Pump 3 9 Q Quit And Save Changes B 9 Quit Without Saving Changes B 9 R Range 3 4 3 10 ranges 1 1 Ranges of Operation 2 1 Rechargeable Battery Pack 2 9 4 1 Rech...

Страница 100: ...hing Off 3 4 3 9 Switching Off the Instrument Pump 3 9 Switching On in CGI Mode 3 8 Switching On in CSM Mode 3 2 System Sampling A 2 T Threshold Display Options 2 4 Time Response A 2 Toxic Ranges 2 5...

Страница 101: ......

Страница 102: ...gmiuk com USA Head Office Detcon 4055 Technology Forest Blvd Suite 100 The Woodlands TX 77381 USA Telephone 713 559 9290 Toll Free 888 367 4286 Fax 281 292 2860 e mail sales detcon com GMI Service Cal...