background image

 

D1062 

- Vibration Transducer Interface 

G.M. International ISM0094-7 

General Description: 

The single channel DIN Rail Vibration Transducer Interface D1062S provides a fully floating dc supply for energizing vibration transducers, accelerometers or 

2-3 wires sensors located in Hazardous Area, and repeats the sensor input voltage in a totally isolated circuit located in Safe Area to drive vibration monitors or analyzers for rotating 
machinery control and supervision purposes. 

Function: 

1 channel I.S. analog voltage input for 2-3 wires sensors, provides 3 port isolation (input/output/supply) and voltage output signal. 

Signalling LED: 

Power supply indication (green). 

Analog Signal Frequency Band: 

DC to 10 KHz within 0.1 dB, 10 KHz to 20 KHz within 3 dB. 

EMC: 

Fully compliant with CE marking applicable requirements. 

Supply: 

24 Vdc nom (20 to 30 Vdc) reverse polarity protected, ripple within voltage limits 

 5 Vpp. 

Current consumption @ 24 V:

 60 mA with 20 mA transducer consumption and 2 mA output load, typical. 

Power dissipation:

 1.1 W with 24 V supply voltage, 20 mA transducer consumption and 2 mA output load typical. 

Max. power consumption:

 at 30 V supply voltage and short circuit condition, 1.9 W. 

Isolation (Test Voltage): 

I.S. In/Out 1.5 KV; I.S. In/Supply 1.5 KV; Out/Supply 500 V. 

Input: 

0 V to -20 V (10 K

 impedance at terminals 15-16). 

3 wires sensor supply voltage: 

more than -21.0 V at 0 mA supply, more than -15.0 V at 20 mA supply (current limited at 

 25 mA) at terminals 14-16. 

2 wires sensor supply voltage: 

more than -15.0 V with constant current supply mode at terminals 15-16. 

Supply current selectable at 4 mA, 6 mA or 10 mA (current supply mode enabled connecting terminals 13-14, 4 mA with terminal 10 not connected, 
6 mA connecting terminals 10-11 or 10 mA connecting terminals 10-12). 

Output: 

0 to -20 V on 10 K

 load, with 10 

 output resistance. 

Response time:

 10 

s (10 to 90 % step change). 

Output ripple:

 

 20 mVrms on 0.5 to 20 KHz band. 

Frequency response:

 DC to 10 KHz within 0.1 dB, 10 KHz to 20 KHz within 3 dB. 

Performance: 

Ref. Conditions 24 V supply, 10 K

 load, 23 ± 1 °C ambient temperature. 

Calibration accuracy: 

 ± 0.05 % of full scale. 

Linearity error: 

 ± 0.05 % of full scale. 

Supply voltage influence: 

 ± 0.005 % of full scale for a min to max supply change. 

Temperature influence:

 

 ± 0.005 % on zero and span for a 1 °C change. 

Compatibility: 

 

CE mark compliant, conforms to 94/9/EC Atex Directive and to 2004/108/CE EMC Directive. 

Environmental conditions:  

Operating:

 temperature limits -20 to + 60 °C, relative humidity max 95 %. 

Storage:

 temperature limits – 45 to + 80 °C. 

Safety Description: 

II (1) G [Ex ia Ga] IIC, II (1) D [Ex ia Da] IIIC, I (M1) [Ex ia Ma] I, II 3G Ex nA II T4, [Ex ia Ga] IIC, [Ex ia Da] IIIC, [Ex ia Ma] I associated electrical apparatus. 
Uo/Voc = 25.9 V, Io/Isc = 90 mA, Po/Po = 576 mW at terminals 14-16. Uo/Voc = 1.1 V, Io/Isc = 12 

A, Po/Po = 4 

W at terminals 15-16; 

Ui/Vmax = 30 V, Ci = 0 nF, Li = 1.5 

H at terminals 15-16 (when used with 3 wires transducer or 2 wires AC sensor connecting terminals 9-14). 

Uo/Voc = 27 V, Io/Isc = 90 mA, Po/Po = 576 mW at terminals 15-16 (when used with 2 wire constant current supply mode connecting terminals 13-14). 
Um = 250 Vrms, -20 °C 

 Ta 

 60 °C. 

Approvals: 

DMT 01 ATEX E 042 X conforms to EN60079-0, EN60079-11, EN60079-26, EN61241-0, EN61241-11, 

IECEx BVS 07.0027X conforms to IEC60079-0, IEC60079-11, IEC60079-26, IEC61241-0, IEC61241-11, IMQ 09 ATEX 013 X conforms to EN60079-0, EN60079-15, 
FM & FM-C No. 3024643, 3029921C, conforms to Class 3600, 3610, 3611, 3810 and C22.2 No.142, C22.2 No.157, C22.2 No.213, E60079-0, E60079-11, E60079-15. 
DNV No.A-13778 Certificates for maritime applications. 

Mounting: 

T35 DIN Rail according to EN50022. 

Weight:

 about 150 g. 

Connection:

 by polarized plug-in disconnect screw terminal blocks to accommodate terminations up to 2.5 mm

2

Location:

 Safe Area/Non Hazardous Locations or Zone 2, Group IIC T4, Class I, Division 2, Groups A, B, C, D Temperature Code T4 and Class I, Zone 2, Group IIC, IIB, IIA T4 installation. 

Protection class:

 IP 20. 

Dimensions:

 Width 22.5 mm, Depth 99 mm, Height 114.5 mm. 

 

 

 

Technical Data 

Characteristics 

Содержание D1062S

Страница 1: ...D1062 Vibration Transducer Interface ISM0094 7 D1062S INSTRUCTION MANUAL Vibration Transducer Interface Din Rail Model D1062S...

Страница 2: ...mVrms on 0 5 to 20 KHz band Frequency response DC to 10 KHz within 0 1 dB 10 KHz to 20 KHz within 3 dB Performance Ref Conditions 24 V supply 10 K load 23 1 C ambient temperature Calibration accuracy...

Страница 3: ...Rail and plug in terminal blocks 250 Vrms Um max voltage allowed to the instruments associated with the barrier For AC Output Sensor if connected to terminal 14 Terminal block connections HAZARDOUS AR...

Страница 4: ...ted 14 16 15 16 with terminals 13 14 connected Li Ri device and L cable R cable IIC A B Lo Ro 61 7 H Lo Ro 247 1 H Lo Ro 494 3 H IIB C IIA D IIC A B Lo Ro 56 8 H Lo Ro 227 3 H Lo Ro 454 7 H IIB C IIA...

Страница 5: ...nsically Safe devices approved with entity parameters not specifically examined in combination as a system when the above conditions are respected For Division 1 and Zone 0 installations the configura...

Страница 6: ...Out 1 11 12 10 9 1 2 In 1 Signal Common Vibration Monitor Programmable Current Source 2 wires Vibration Transducer AC Signal Common AC Output Sensor Jumper Sensor operating in AC signal mode Jumpers...

Страница 7: ...ning substitution of components may impair Intrinsic Safety and suitability for Division 2 Zone 2 Explosion Hazard to prevent ignition of flammable or combustible atmospheres disconnect power before s...

Страница 8: ...milar sensor connect sensor negative supply wire at terminal 14 sensor signal wire at terminal 15 and sensor common wire identical for both signal and supply at terminal 16 For 2 wires powered sensor...

Отзывы: