Rhinomanometer NR6
NR6 User
’s
Manual V 16 14/02/2020 P26 of 34
SERVICING
Servicing can only be carried out by GMI approved and
authorised personnel. No modifications are allowed
.
The NR6 contains two transducers with associated instrumentation amplifiers, balance and gain
controls, and an is5 Volts to +8V DC converter. Full circuit diagrams are available on
request and service adjustments are noted below.
TRANSDUCER SETUP ADJUSTMENTS
1. Switch on the NR6 and allow 5 minutes to warm up.
2. Measure the voltage between test point 5 and test point 4 using a sensitive DC voltmeter.
Adjust pressure offset potentiometer VR4 to give a reading of 0 volts.
3. Measure the voltage between test point 5 and test point 6 using a sensitive DC voltmeter.
Adjust flow offset potentiometer VR1 to give a reading of 0 volts.
TP5
TP4
TP6
VR1
VR4