Important Control Solution Information
Use only
TaiDoc
control solutions with your meter.
Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution vial
and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature (20°C-25°C / 68°F-77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
2°C and 30°C (35.6°F and 86°F). Do NOT freeze.
SYSTEM TROUBLESHOOTING
If you follow the recommended action but the problem persists, or
error messages other than the ones below appear, please call your
local customer service. Do not attempt to repair by yourself and
never try to disassemble the meter under any circumstances.
Appears
When Glucose
<1.1mmol/L (20mg/dL)
≧
15mmol/L (270mmol/L)
>33.3mmol/L (600mg/dL)
Result Readings