Preparation: 

 

a.

 

Set up a test stand with a rod and clamp to hold the sensor. Beakers 
will be placed underneath the clamp so the sensor can be inserted into 

the liquid.   

b.

 

The sensor must have the stainless steel section of the body in contact 
with the test solution.  Install the sensor so that the end of the sensor 
is at least 3 inches from the bottom.  If the sensor is too close to the 

bottom, infrared light may be reflected from the glass and effect the 
measurements. 

c.

 

The lenses must be gently cleaned with a cotton swab and isopropyl 
alcohol before each calibration step. 

d.

 

Fill a large beaker with distilled water.  This will be used to rinse 
solution from the sensor in between steps. 

e.

 

Fill three more beakers with 0.0 NTU Standard, 50 NTU Standard and 
1000 NTU Standard.  These beakers must be at least 500 ml in size or 

the end of the sensor will be too close to the bottom for accurate 
calibration.  This generally requires solution volumes of at least 500ml. 

f.

 

The calibration process should not be done in bright light if possible.  
While not generally required, the extra step of covering the outside of 

the beakers with a non-transparent material may help to prevent stray 
light from the outside affecting the readings. 

 

Note:  The sensor is factory calibrated and will provide output currents 

of 20mA at the maximum turbidity values; however, some small drift in 
this output current can occur over time.  These adjustments are done 
internally, access to these settings are not available to the user. The 
sensors should never be disassembled.  The following process 

describes how to re-check these output currents.  A full recalibration 
can only be done by returning the sensor to Global Water. 

 
 
 

Содержание WQ730

Страница 1: ...Global Water 800 876 1172 globalw com 1 Global Water Instrumentation Inc 11390 Amalgam Way Gold River CA 95670 T 800 876 1172 Int l 916 638 3429 F 916 638 3270 Turbidity Sensor WQ730 02 002 11 30 09...

Страница 2: ...ill find this product to be a valuable asset for your application Should you require assistance our technical staff will be happy to help Table of Contents I Use and Maintenance Page 3 II Sensor Speci...

Страница 3: ...asive cleaners on the lenses or permanent damage will occur Cleaner water requires cleaning less often in other applications the sensor may need to be cleaned once each day d The calibration of the se...

Страница 4: ...8 seconds recommended Operating Temperature 0 C to 50 C Storage Temperature 10 C to 60 C Dimensions 1 1 2 diameter x 7 1 2 long Weight 1 lb plus cable b Connections Red 10 36 VDC Black Ground Green 4...

Страница 5: ...com f In the event that the equipment needs to be returned to the factory for any reason please call to obtain a RMA Return Material Authorization Do not return items without a RMA displayed on the ou...

Страница 6: ...obligations under this warranty are limited to at Global Water s option I replacing or II repairing any products determined to be defective In no case shall Global Water s liability exceed the produc...

Страница 7: ...gle of 90 degrees to the incident light beam measuring side scattered light The detected light intensity is directly proportional to the turbidity of the water b The following supplies are required St...

Страница 8: ...TU Standard 50 NTU Standard and 1000 NTU Standard These beakers must be at least 500 ml in size or the end of the sensor will be too close to the bottom for accurate calibration This generally require...

Страница 9: ...U standard Turn on the power supply and the current meter Let the sensor stabilize for several seconds before taking any measurements If necessary use a lint free cotton swab bent at 90 to remove any...

Страница 10: ...ndard Solutions The following turbidity standards are available from Global Water contact the Sales Department or Technical Support for more information These are EPA approved primary standards that a...

Страница 11: ...Global Water 800 876 1172 globalw com 11 Appendix C Sensor Measurement Diagram...

Страница 12: ...al equipment operating within certain voltage ranges According to the following standards EN61010 1 2001 Electro Magnetic Directives 2004 108 EC Requirements to prevent electrical and electronic equip...

Отзывы: