Appendix A:  Turbidity Sensor Calibration 

 
Note: The following steps show how to calibrate the WQ730 Turbidity 
Sensor using suspended polymer turbidity standards only. 

 

General 
 

a.

 

The Turbidity Sensor is measuring turbidity in liquids by using a 90 

degrees scatter nephelometer where a focused beam is directed into 
the monitored water. The principle is based on recommendations of 
the international guideline ISO

 

7027

.  

Two detectors receive light from 

a source emitting 880 nm infrared light in a pulsed mode.  One 

detector is placed opposite the source, receiving the transmitted light, 
and the other detector is placed at an angle of 90 degrees to the 
incident light beam measuring side-scattered light.  The detected light 

intensity is directly proportional to the turbidity of the water. 

b.

 

The following supplies are required: 
Stand with rod and clamp 
Beakers, 500 or 1000 ml 

1000 NTU Turbidity Standard 
50 NTU Turbidity Standard 
0.0 NTU Turbidity Standard 
Distilled Water 

Power Supply 
Current Meter 
Test leads as necessary 
 

The preferred and recommended turbidity standards are suspended 
polymer solutions.  These are EPA approved, non-toxic, and require 
no mixing or stirring.  Other types of standards such as Formazin can 
also be used, but these require thorough mixing and must be used 

with low speed stirring plates during the calibration process. 
 
All the necessary turbidity standard solutions are available from Global 
Water, contact the sales department for more information.  See 

Appendix B. 

 

Содержание WQ730

Страница 1: ...Global Water 800 876 1172 globalw com 1 Global Water Instrumentation Inc 11390 Amalgam Way Gold River CA 95670 T 800 876 1172 Int l 916 638 3429 F 916 638 3270 Turbidity Sensor WQ730 02 002 11 30 09...

Страница 2: ...ill find this product to be a valuable asset for your application Should you require assistance our technical staff will be happy to help Table of Contents I Use and Maintenance Page 3 II Sensor Speci...

Страница 3: ...asive cleaners on the lenses or permanent damage will occur Cleaner water requires cleaning less often in other applications the sensor may need to be cleaned once each day d The calibration of the se...

Страница 4: ...8 seconds recommended Operating Temperature 0 C to 50 C Storage Temperature 10 C to 60 C Dimensions 1 1 2 diameter x 7 1 2 long Weight 1 lb plus cable b Connections Red 10 36 VDC Black Ground Green 4...

Страница 5: ...com f In the event that the equipment needs to be returned to the factory for any reason please call to obtain a RMA Return Material Authorization Do not return items without a RMA displayed on the ou...

Страница 6: ...obligations under this warranty are limited to at Global Water s option I replacing or II repairing any products determined to be defective In no case shall Global Water s liability exceed the produc...

Страница 7: ...gle of 90 degrees to the incident light beam measuring side scattered light The detected light intensity is directly proportional to the turbidity of the water b The following supplies are required St...

Страница 8: ...TU Standard 50 NTU Standard and 1000 NTU Standard These beakers must be at least 500 ml in size or the end of the sensor will be too close to the bottom for accurate calibration This generally require...

Страница 9: ...U standard Turn on the power supply and the current meter Let the sensor stabilize for several seconds before taking any measurements If necessary use a lint free cotton swab bent at 90 to remove any...

Страница 10: ...ndard Solutions The following turbidity standards are available from Global Water contact the Sales Department or Technical Support for more information These are EPA approved primary standards that a...

Страница 11: ...Global Water 800 876 1172 globalw com 11 Appendix C Sensor Measurement Diagram...

Страница 12: ...al equipment operating within certain voltage ranges According to the following standards EN61010 1 2001 Electro Magnetic Directives 2004 108 EC Requirements to prevent electrical and electronic equip...

Отзывы: