Global Leak Detection GLD-40 Скачать руководство пользователя страница 13

9

Troubleshooting Guide

Your Smoke Wizard has two indicator lights on the control panel that double as diagnostic lights,

indicating if the Tester is working correctly. The following table describes the indicator light function
codes.

Green

Red

Interval

Cause

Constant ON

Sufficient battery power

Blinks: 1 per second

Insufficient battery power

Blink simultaneously: 1 per second

Bad ground or power connection at smoke
canister or short in circuit

Blink simultaneously: 4 times per second

Bad ground at smoke canister or open
heating circuit

Blink alternately: 1 per second
(System will shut down; Power must be
disconnected then reconnected)

Bad ground. Can also mean circuit board
failure

٭

Symptom

Likely Cause

Solution

The green power
indicator lamp on the
Tester does not turn
ON.

1. The power cables are reversed.

2. Poor power-supply cable
connection.

3. Battery providing power is too
weak.

1. Correctly position the power cables.

2. Secure the connection at the positive
terminal and chassis ground.

3. Verify the battery is in good condition
and fully charged.

I turn the Tester ON but
there is no air or smoke
coming out of the
supply hose.

1. Flow Control valve is closed.

2. Poor power-supply cable
connection.

3. Battery providing power is too
weak.

4. Air supply to tester is
insufficient.

5. Smoke-vapor could be
condensing inside the smoke
supply hose, restricting smoke
flow.

1. Open flow control.

2. Secure the connection at the positive
terminal and chassis ground.

3. Verify the battery is in good condition
and fully charged.

4. Be sure air supply is sufficient.

5.

Be sure the smoke hose is fully

extended when in use. Drain the
condensation from the hose prior to its
next use, either by placing the machine
higher than the hose, or by removing the
hose and blowing through it with
compressed air. Do NOT blow
compressed air into the machine with the
smoke hose connected to the machine.

٭

If circuit board failure occurs, first try disconnecting power to your Smoke Wizard for 10 seconds and

reconnect. If this failure code occurs a second time, disconnect Smoke Wizard and contact Smoke Wizard UK
at: [email protected].

Содержание GLD-40

Страница 1: ...OPERATOR S MANUAL Diagnostic Smoke Vapor Machine with UltraTraceUV Dye Solution Leak Detection System Model No GLD 40 USA Canada Patents 6 392 227 6 526 808 2 279 147 Int l Patents Pending...

Страница 2: ......

Страница 3: ...be between 50 175 PSI 3 4 to 12 bar Connect to workshop airline for general purpose leak detection applications Smoke exiting a very small leak is sometimes even easier to see if after filling the sy...

Страница 4: ...e leak detection smoke machine available today The patented technology inside your Smoke Wizard was designed in collaboration with major automakers was designed specifically to be safe for leak testin...

Страница 5: ...remove Cap Plugs Kit GLD005 these cap plugs are useful when sealing an intake manifold system in order to contain the smoke in the system for proper leak testing such as mass airflow sensor ducting e...

Страница 6: ...end it The patented UltraTraceUV smoke solution supplied with this Tester will perform hundreds of tests that only cost pennies per test is the only solution in the world approved by the auto manufact...

Страница 7: ...Smoke Wizard from a convenient test location on the vehicle or for storage when not in use Note Always use bonnet support device when hanging from the bonnet Filter Drain Port located under Tester ke...

Страница 8: ...the leak is even easier to see Sometimes less smoke exiting a small leak is easier to see Flow Control Valve does not affect delivery pressure it only affects the volume of flow SMOKE Light Red Turns...

Страница 9: ...ter 5 Connect smoke supply hose to system being tested 6 Flow Control Valve must be in Full Open position Figure 3 7 Press START button and fill system with smoke Whenever possible it is best to purge...

Страница 10: ...aving a leak such as intake induction system intercooler and turbocharger system vacuum system exhaust system wind water leaks and can also be used to verify air solenoid functions and test components...

Страница 11: ...iv ve e c ca ab bi in n p pr re es ss su ur re e 2 2 C Co on nn ne ec ct t T Te es st te er r s s s su up pp pl ly y h ho os se e n no oz zz zl le e t to o S Sm mo ok ke e D Di if ff fu us se er r 3 3...

Страница 12: ...location of a leak Use the UV light provided to highlight the dye This technology is designed so that the dye will only deposit if there is pressure differential So for instance the dye will deposit w...

Страница 13: ...itive terminal and chassis ground 3 Verify the battery is in good condition and fully charged I turn the Tester ON but there is no air or smoke coming out of the supply hose 1 Flow Control valve is cl...

Страница 14: ...ormal use care and service the Equipment except as otherwise provided herein shall be free from defects in material and workmanship for ONE YEAR from the date of original invoice Seller s obligations...

Страница 15: ...ry injury may occur EMERGENCY OVERVIEW INHALATION INGESTION The balance of the ingredients are proprietrary and non hazardous as defined in US 29CFR 1210 and European regulation directive 67 548 EEC S...

Страница 16: ...R PROTECTIVE CLOTHING OR EQUIPMENT HAZARDOUS COMBUSTION PRODUCTS Carbon Monoxide Soot Smoke Carbon Dioxide Non Hazardous SECTION 4 FIRST AID MEASURES SECTION 5 FIRE FIGHTING MEASURES Do not induce vom...

Страница 17: ...NA NA NA UN NA IDENTIFICATION NUMBER NA SECTION 14 TRANSPORT INFORMATION Not regulated Not regulated IATA HAZARD CLASS Pack Group SECTION 12 ECOLOGICAL INFORMATION No ecological information is availa...

Страница 18: ...so ChemTel Inc assumes no responsibility for document alterations or additions after authorship and initial review excepting changes performed by ChemTel personnel Data and calculations are based on...

Страница 19: ...is report by itself does not imply that the material product or service is or has ever been under an Intertek certification program Intertek Testing Services NA Inc EMC VERIFICATION No 3179331LAX 001g...

Страница 20: ......

Страница 21: ...ry injury may occur EMERGENCY OVERVIEW INHALATION INGESTION The balance of the ingredients are proprietrary and non hazardous as defined in US 29CFR 1210 and European regulation directive 67 548 EEC S...

Страница 22: ...R PROTECTIVE CLOTHING OR EQUIPMENT HAZARDOUS COMBUSTION PRODUCTS Carbon Monoxide Soot Smoke Carbon Dioxide Non Hazardous SECTION 4 FIRST AID MEASURES SECTION 5 FIRE FIGHTING MEASURES Do not induce vom...

Страница 23: ...NA NA NA UN NA IDENTIFICATION NUMBER NA SECTION 14 TRANSPORT INFORMATION Not regulated Not regulated IATA HAZARD CLASS Pack Group SECTION 12 ECOLOGICAL INFORMATION No ecological information is availa...

Страница 24: ...so ChemTel Inc assumes no responsibility for document alterations or additions after authorship and initial review excepting changes performed by ChemTel personnel Data and calculations are based on...

Страница 25: ...is report by itself does not imply that the material product or service is or has ever been under an Intertek certification program Intertek Testing Services NA Inc EMC VERIFICATION No 3179331LAX 001g...

Отзывы: