Model GLI-93711 Portable Chlorine Meter

solve this, neutralize the sample with diluted
HCl or NaOH.

In case of water with hardness greater than
500 mg/l CaCO

3

, shake the sample for about

1 minute after adding the reagent.

TIPS FOR AN ACCURATE

MEASUREMENT

The tips listed below should be carefully fol-
lowed during testing to ensure best accuracy:

 Do not touch the cuvet walls with hands.

 To maintain the same conditions during

the zeroing and measuring phases, it is
necessary to close the cuvet to prevent
any contamination.

 Do not let test sample stand too long after

reagent is added or accuracy will be lost.

 Whenever the cuvet is placed into the

measurement cell, it must be completely
free of fingerprints, oil or dirt. Wipe it thor-
oughly with a lint-free cloth prior to
insertion.

 It is important that the sample does not

contain any debris. This would corrupt the
readings.

 It is possible to take multiple readings in a

row, but it is recommended that a zero
reading be taken for each sample and that
the same cuvet is used for zeroing and
measurement.

 It is important to discard the sample im-

mediately after the reading is taken
because the glass might become perma-
nently stained.

 Shaking the cuvet can generate bubbles

in the sample, causing higher readings.
To obtain accurate measurements, re-
move such bubbles by swirling or by
gently tapping the vial.

All the reaction times reported in this manual
are referred to 20°C (68°F). As a general

rule, they should be doubled at 10°C (50°F)
and halved at 30°C (86°F).

BATTERY REPLACEMENT

When the battery is low,
"BA" is displayed to indicate
that it must be replaced.

Battery replacement must only take place in a
non-hazardous area. To replace a run-down
battery, simply remove the battery compart-
ment cover on the rear of the meter and
replace the battery, observing correct polarity.

WARRANTY

GLI International, Inc. warrants the Model
GLI-93711 Hand-held/Portable Chlorine Me-
ter to be free from defects in material or
workmanship for a period of one year (12
months) from the date of shipment of this
product from our facility. A warranty claim will
not be honored if defects are not reported
within the warranty period, or if GLI Interna-
tional determines that defects or damages
are due to normal wear, misapplication, lack
of maintenance, abuse, improper installation,
alteration or abnormal conditions. GLI Inter-
national’s obligation under this warranty shall
be limited to, at its option, replacement or re-
pair of this product. The product must be
returned to GLI International, freight prepaid,
for examination. The product must be thor-
oughly cleaned and any process chemicals
removed before it will be accepted for re-
placement or repair. GLI International’s
liability shall not exceed the cost of the prod-
uct. Under no circumstances will GLI
International be liable for any incidental or
consequential damages, whether to person
or property. GLI International will not be liable
for any other loss, damage or expense of any
kind, including loss of profits, resulting from
the installation, use, or inability to use this
product.

Содержание GLI-93711

Страница 1: ...Manual No 93711 Revision 0 401 MODEL GLI 93711 HAND HELD PORTABLE TOTAL FREE CHLORINE METER ...

Страница 2: ...age notify the transit carrier immediately Each meter is supplied complete with Two sample cuvets and caps One transport cap Instruction Manual One 9 Volt Battery NOTE Save all packing material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The meter measures the...

Страница 3: ...he meter is in a ready state and ze roing can be performed Sampling in Progress This prompt appears each time the meter is performing a measurement This indicates that the meter is in a zeroed state and measurement can be per formed A zero reading was not taken Insert a sample be fore adding reagent and press ZERO Under range A blinking 0 00 indicates that the sample absorbs less light than the ze...

Страница 4: ...Free Chlorine add 3 drops of DPD1 indicator Bottle A and 3 drops of DPD1 buffer Bottle B to the other cuvet To measure Total Chlorine also add 1 drop of DPD3 solution Bottle C Shake gently before adding 10 ml of unreacted sample from the first cuvet Replace the cap and shake gently again 10 Reinsert the cuvet into the instrument 11 Wait for approximately 30 seconds and then press READ The display ...

Страница 5: ... 68 F As a general rule they should be doubled at 10 C 50 F and halved at 30 C 86 F BATTERY REPLACEMENT When the battery is low BA is displayed to indicate that it must be replaced Battery replacement must only take place in a non hazardous area To replace a run down battery simply remove the battery compart ment cover on the rear of the meter and replace the battery observing correct polarity WAR...

Отзывы: