background image

Model GLI-93711 Portable Chlorine Meter

4

Rev. 0-401

OPERATIONAL GUIDE

MEASUREMENT PROCEDURE

1.

Turn the meter on by press-
ing ON/OFF.

2.

When the LCD displays
"- - -", it is ready.

3.

Select free or total Chlorine by pressing
FREE/TOTAL. In the lower right corner,
an "I" will appear to indicate free chlorine
or a "C" will appear to indicate total.

4.

Fill the cuvet with 10 ml of
unreacted sample, up to
the mark, and replace the
cap.

5.

Place the cuvet into
the holder and ensure
that the notch on the
cap is positioned se-
curely into the groove.

6.

Press ZERO and "SIP" will appear on the
display.

7.

Wait for a few seconds and the display
will show "-0.0-". Now the
meter is zeroed and
ready for measurement.

8.

Remove the cuvet.

9.

To measure 

Free Chlorine

, add 3 drops

of DPD1 indicator (“Bottle A”) and 3

drops of DPD1 buffer (“Bottle B”) to the
other cuvet. To measure 

Total Chlorine

,

also add 1 drop of DPD3 solution (“Bottle
C”). Shake gently before adding 10 ml of
unreacted sample from the first cuvet.
Replace the cap and shake gently again.

10. Reinsert the cuvet into the instrument.

11. Wait for approximately 30 seconds and

then press READ. The display will show
"SIP" during measurement.

The instrument directly displays concentra-
tion in mg/l of free or total chlorine.

NOTE:

Free and total chlorine have to be
measured separately with fresh un-
reacted samples following the
previously described procedure if
both values are needed.

INTERFERENCES

Interference may be caused by:

 Bromine

 Iodine

 Fluorine

 Ozone

 Oxidized manganese

 Chromium

Alkalinity above 250 mg/l or acidity above
150 mg/l will not reliably develop the full
amount of color or it may rapidly fade. To re-

Содержание GLI-93711

Страница 1: ...Manual No 93711 Revision 0 401 MODEL GLI 93711 HAND HELD PORTABLE TOTAL FREE CHLORINE METER ...

Страница 2: ...age notify the transit carrier immediately Each meter is supplied complete with Two sample cuvets and caps One transport cap Instruction Manual One 9 Volt Battery NOTE Save all packing material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The meter measures the...

Страница 3: ...he meter is in a ready state and ze roing can be performed Sampling in Progress This prompt appears each time the meter is performing a measurement This indicates that the meter is in a zeroed state and measurement can be per formed A zero reading was not taken Insert a sample be fore adding reagent and press ZERO Under range A blinking 0 00 indicates that the sample absorbs less light than the ze...

Страница 4: ...Free Chlorine add 3 drops of DPD1 indicator Bottle A and 3 drops of DPD1 buffer Bottle B to the other cuvet To measure Total Chlorine also add 1 drop of DPD3 solution Bottle C Shake gently before adding 10 ml of unreacted sample from the first cuvet Replace the cap and shake gently again 10 Reinsert the cuvet into the instrument 11 Wait for approximately 30 seconds and then press READ The display ...

Страница 5: ... 68 F As a general rule they should be doubled at 10 C 50 F and halved at 30 C 86 F BATTERY REPLACEMENT When the battery is low BA is displayed to indicate that it must be replaced Battery replacement must only take place in a non hazardous area To replace a run down battery simply remove the battery compart ment cover on the rear of the meter and replace the battery observing correct polarity WAR...

Отзывы: