background image

Model GLI-9143 Portable D.O. Meter

6

Rev. 0-401

For an accurate measurement, allow suffi-
cient time for thermal equilibrium between the
probe and the measurement sample (a few
minutes if the temperature difference is sev-
eral degrees).

ppm READINGS

The ppm readings make it possible to read
the concentration of the dissolved oxygen di-
rectly.

If the sample contains significant salinity or if
the measurement is taken at a higher altitude
than sea level, the readout values must be
corrected (by taking into account the lower
degree of oxygen solubility in such conditions
-- see below).

Set the altitude and/or the salinity before cali-
bration and taking ppm measurements. The
meter will automatically compensate for these
factors.

% O

2

 SATURATION READINGS

The % O

2

 saturation reading provides the

rate of oxygen saturation with reference to
100.0% at sea level.

TEMPERATURE READINGS

The lower part of the dis-
play will show the
measured temperature in
degrees Celsius.

Allow the probe to reach thermal equilibrium
with the sample before taking any measure-
ment. The greater the difference between the
ambient temperature and the temperature of
the sample, the longer it will take the probe to
equilibrate to the sample.

ALTITUDE COMPENSATION

Press FACTOR and “F1”
will be displayed. Use the
UP and the DOWN keys to
set the altitude from 0-4000
m (0-13,123 ft.), in steps of
100 m (328 ft.).

SALINITY COMPENSATION

Press FACTOR twice and
“F2” will be displayed.
Press UP and DOWN to
set the salinity between 0
and 80 g/l. Press FACTOR
again to display the tem-
perature.

PROBE AND MEMBRANE MAINTENANCE

The dissolved oxygen probe body is made of
reinforced polypropylene for maximum dura-
bility. A built-in thermistor measures the
temperature of the sample. It is recom-
mended that the protective cap always be
kept on the probe when the probe is not in
use.

To replace the membrane cap 

or refill it with

electrolyte, proceed as follows:

1.

Remove the protective
cap by gently twisting
and pulling it off the
probe (see Figure 1).

FIGURE 1

Содержание GLI-9143

Страница 1: ...Manual No 9143 Revision 0 401 MODEL GLI 9143 HAND HELD PORTABLE DISSOLVED OXYGEN METER ...

Страница 2: ... material until you are sure that the instrument func tions correctly All defective items must be returned in their original packaging together with the supplied accessories GENERAL DESCRIPTION The GLI 9143 is a water resistant micro processor based auto calibration Dissolved Oxygen meter with automatic temperature compensation ATC It has been developed for dissolved oxygen and temperature measure...

Страница 3: ... Polypropylene Probe Body 5 Temperature Sensor 6 O ring Seal 7 Silver Chloride Anode sensor element 8 Platinum Cathode sensor element 9 Oxygen permeable Teflon Membrane 10 Protective Cap FUNCTIONAL DESCRIPTION METER 1 Battery compartment 2 Probe connector 3 Liquid crystal display 4 ON OFF button 5 CAL button to enter or exit calibration mode 6 FACTOR button to select altitude F1 and salinity F2 7 ...

Страница 4: ...76407A P supplied with the meter and make sure that the rubber O ring sits properly inside the membrane cap 4 Rinse the mem brane cap with electrolyte while shaking it gently Refill with clean electrolyte 5 Gently tap the sides of the mem brane cap with your finger tip to ensure that no air bubbles remain trapped To avoid damaging the membrane do not tap the membrane cap directly on the bottom 6 W...

Страница 5: ...ss FACTOR and ensure F1 and F2 are set to the appropriate altitude and salinity values see ALTITUDE and SALINITY COMPENSATION NOTE The instrument must be calibrated whenever the probe membrane or the electrolyte is changed To exit the calibration mode during calibration press CAL The display may be switched from readings in saturation to ppm without re calibration by simply pressing RANGE MEASUREM...

Страница 6: ...y will show the measured temperature in degrees Celsius Allow the probe to reach thermal equilibrium with the sample before taking any measure ment The greater the difference between the ambient temperature and the temperature of the sample the longer it will take the probe to equilibrate to the sample ALTITUDE COMPENSATION Press FACTOR and F1 will be displayed Use the UP and the DOWN keys to set ...

Страница 7: ...shed or stained which could be due to contact with certain gases or extended use with a loose or damaged mem brane the cathode should be cleaned 1 Use a clean lint free cardboard or cloth and rub the cathode very gently side to side 4 5 times This will be enough to polish and remove any stains without damaging the platinum tip 2 Rinse the probe with deionized or dis tilled water 3 Install a new me...

Страница 8: ...Meter to be free from defects in material or workmanship for a period of one year 12 months from the date of shipment of this product from our facility The GLI 76407 4 or GLI 76407 10 Dissolved Oxygen Sensor is warranted for a period of 6 months A war ranty claim will not be honored if defects are not reported within the warranty period or if GLI International determines that defects or damages ar...

Отзывы: